16060-101 Rev. M
39
General Specifications
Model:
16060 - Reichert 7CR
Physical Dimensions
Size:
Weight, unpacked: 23.0 lbs. (10.4 Kg)
Height: 19.8 in. (50.2 cm)
Width: 10.5 in. (26.7 cm)
Depth: 14.0 in. (35.6 cm)
Electrical
Voltage: 100-240 VAC
Power: 60-85 VA
Frequency: 50/60 Hz
Fuses: Time-Lag (2.5A, 250V), 5 X 20mm, RoHS
Measurement Range:
7 – 60 mmHg
Measurement Accuracy:
Accuracy by NIST traceable pressure transducer (95% Confidence)
IOPg:
± 1.0 mmHg
(7 - 60 mmHg)
IOPcc:
± 1.0 mmHg
(7 - 60 mmHg)
Operational Conditions
Environmental:
The environmental conditions are as follows:
Operating:
Temperature:
10° C (50° F) to 35° C (95° F)
Relative Humidity: 30% to 90%
Atmospheric Pressure: 80
(23.6 in. Hg)
to
106 kPa
(31.3 in. Hg)
Transportation & Storage:
Temperature:
-40° C (-40° F) to +70° C (158° F)
Relative Humidity: 10% to 95%
Atmospheric Pressure: 50
(14.8 in. Hg)
to
106 kPa
(31.3 in. Hg)
Disposal
This product does not generate any environmentally hazardous residues. At the end of its product
service life, follow your local laws and ordinances regarding the proper disposal of this equipment.
Software Revision
The software revision can be obtained by contacting Reichert Technologies. The serial number
identifies the manufacture date and will provide access to the software version.
-40°C
70°C
10%
95%
106 kPa
50 kPa