Guidance Tables
Table 201 – Guidance and Manufacturer’s Declaration
Electromagnetic Emissions
All Medical Electrical Equipment and Medical Electrical Systems
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
The Reichert 7CR is intended for use in the electromagnetic environment specified below. The
customer or user of the Reichert 7CR should ensure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic Environment
- Guidance -
Conducted and Radiated
RF Emissions
CISPR 11
Group 1
The Reichert 7CR uses RF energy only for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby
electronic equipment.
Conducted and Radiated
RF Emissions
CISPR 11
Class A
The Reichert 7CR is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply network
that supplies building for domestic power.*
Harmonic Distortion
IEC 61000-3-2
Class A
Voltage Fluctuations and
Flicker
IEC 61000-3-3
Complies
Note:
*The emissions characteristics of this equipment make it suitable for use in industrial areas and
hospitals (CISPR 11 class A). If it is used in a residential environment (for which CISPR 11 class
B is normally required) this equipment might not offer adequate protection to radio-frequency
communication services. The user might need to take mitigation measures, such as relocating or re-
orienting the equipment.