20
User Guide
EN
Dansensor
®
LeakPointer
H
2
O
4. Troubleshooting
Vacuum cannot be established
Does ejector start when pushing start
button?
NO
Check that air supply is connected.
Check that no hoses are bent.
Check that hoses are connected
according to flow drawing - see
Check that flow adjustment is fully open
YES
With ejector started and lid open, can
you feel/hear vacuum in air-vent hole?
NO
Check that no hoses are bent.
Check that hoses are connected
according to flow drawing - see
Check that flow adjustment is fully open
Clean ejector and make sure that silencer
is not blocked.
YES
Are sealing without damages and
correct mounted.
NO
Mount sealing correct - see
Replace sealing - see
for ordering
details.
YES
When closing the lid, does sealing seal
properly against lid without air gap?
NO
Adjust lid - see
YES
When closing the lid and starting ejector,
can you feel vacuum from balance hose
inlet?
NO
Remove balance hose
Check that check valve in quick coupler
works. If not replace quick-coupler - see
for ordering details.
YES
Is hose between vacuum gauge and lid
mounted correct?
NO
Mount hose according to flow drawing -
see
"Internal connections" on page 29
.
YES
Call MOCON Europe A/S Support Dept.
Содержание mocon Dansensor LeakPointer H2O
Страница 1: ...P N 390225 C 11 2021 UserGuide...
Страница 2: ...This blank page has been inserted to enable double sided printing of the document...
Страница 4: ...2 User Guide EN Dansensor LeakPointer H2O P N 390225 C 11 2021...
Страница 35: ...This blank page has been inserted to enable double sided printing of the document...
Страница 36: ...0 0 VSPQF 4 3 OOFEFWFK 3JOHTUFE FONBSL 5FM JOGP EL NPDPO BNFUFL DPN XXX BNFUFLNPDPO DPN...