Dansensor
®
LeakPointer
H
2
O
EN
User Guide
17
Test with balance hose
In some cases, there is not enough air in the package to create a pressure inside the package
(DeltaP) high enough to detect a leak.
In these cases, we recommend using the supplied balance hose, which allows air to be sucked
into the package from the outside during the test.
NOTE! The DeltaP in the package will now be the same as the tub vacuum.
1. Make sure that the fitting
is blocked, either by means of the plug
or the pressure
measuring hose
from the digital pressure meter (option - see
2. If you want to use the digital pressure meter (option - see
) to measure the tub
vacuum, this should be connected as described in
"Measuring the tub vacuum" on page 19
3. Place a septum
on the package
and insert the needle
.
CAUTION! Be careful not to insert the needle into the package contents, as this
may soil or block the needle.
4. Connect the hose connector
to the quick coupling
located on the underside of the
lid in the rear left corner.
5. Place the package
in the water and start the test - see
.
CAUTION! The pressure inside the package (DeltaP) will be the same as the vacuum
in the tub, so it is important that the test vacuum is not too high, as this may cause
the package to burst (water must be exchanged).
Depending on the package size it might take some time to balance the pressure
inside the package.
Содержание mocon Dansensor LeakPointer H2O
Страница 1: ...P N 390225 C 11 2021 UserGuide...
Страница 2: ...This blank page has been inserted to enable double sided printing of the document...
Страница 4: ...2 User Guide EN Dansensor LeakPointer H2O P N 390225 C 11 2021...
Страница 35: ...This blank page has been inserted to enable double sided printing of the document...
Страница 36: ...0 0 VSPQF 4 3 OOFEFWFK 3JOHTUFE FONBSL 5FM JOGP EL NPDPO BNFUFL DPN XXX BNFUFLNPDPO DPN...