Remember to use appropriate protective equipment or get help when carrying the
calibrator (for a longer distance) in order to prevent injuries from dropping the
calibrator.
About the front panel:
•
The connectors, on the front panel of the calibrator, must NEVER be connected
to a voltage source.
•
Thermostats connected to the switch test input must not be connected to any
other voltage source during a test.
About the insertion tubes and insulation plug:
•
Never leave hot insertion tubes which have been removed from the calibrator
unsupervised – they may constitute a fire hazard or personal injury.
If you intend to store the calibrator in the optional aluminium carrying case after
use, you must ensure that the instrument has cooled down to a temperature
below 50°C/122°F before placing it in the carrying case.
•
Never place a hot insertion tube in the optional carrying case.
About the fuses:
•
The fuse box must not be removed from the power control switch until the mains
cable has been disconnected.
•
The two main fuses must have the specified current and voltage rating and be of
the specified type. The use of makeshift fuses and the short-circuiting of fuse
holders are prohibited and may cause a hazard.
Caution – Hot surface
This symbol is visible on the grid plate.
•
Do not touch the grid plate, the well or the insertion tube as the calibrator is
heating up – they may be very hot and cause burns.
•
Do not touch the tip of the sensor when it is removed from the insertion
tube/well – it may be very hot and cause burns.
•
Do not touch the handle of the calibrator during use – it may be very hot and
cause burns.
•
Over 50°C/122°F
If the calibrator has been heated up to temperatures above 50°C/122°F, you must
wait until the instrument reaches a temperature below 50°C/122°F before you
switch it off.
•
Do not remove the insert from the calibrator before the insert has cooled down to
less than 50°C/122°F.
Caution…
About the use:
•
Do not use the instrument if the fan is out of order.
•
Before cleaning the calibrator, you must switch it off, allow it to cool down and
remove all cables.