10.0 Errors
Warning
During set-up, make sure that there is correlation between the software set-up, and the
actual distribution of the sensors connected to the calibrator. The software cannot
detect misplacements of sensors.
If the power is lost to the entire test set-up or part of it, it is highly recommended to restart the
calibration.
GENERAL ERRORS
Error description
Solution
ASC-400 does not turn on :
Check the batteries
Check the power supply
If the error still occurs, return the ASC-400 to the manufacturer for service
Faulty readout :
Check that the connected sensors are correctly connected and that they are
properly polarised
Check that the ASC-400 is set up correctly
Switch the ASC-400 off and on again
If the error still occurs, return the ASC-400 to the manufacturer for service
No contact with the ASC-400 :
Check the power supply
Check the USB connection. Use only the supplied cable
Switch the ASC-400 off and on again - and set up the software again
If the error still occurs, return the ASC-400 to the manufacturer for service
When using Hyper Terminal or similar terminal programs, and communications
fails, check that the program has been set up correctly.
Faulty readout from
thermocouples:
Check that the correct type has been selected during setup
Check that the sensors have been correctly polarised and connected
Check that the correct type of cold junction has been selected
manually/automatically/off
Check that the ASC-400 is not subjected to direct or indirect heat sources
Содержание Jofra ASC-400
Страница 1: ...Reference manual Advanced Signal Calibrator Jofra ASC 400...
Страница 2: ...Reference Manual Advanced Signal Calibrator JOFRA ASC 400 Copyright 2014 AMETEK Denmark A S...
Страница 8: ...128797 05 2017 08 11 7 1 3 Dimensioning drawing 55 mm 96 mm 220 mm...
Страница 49: ...128797 05 2017 08 11 48 Figure 19 Calibrating an I P Device APM pressure module Pressure Device...
Страница 52: ...128797 05 2017 08 11 51 Figure 21 Calibrating a Pressure Transmitter Hand pump APM pressure module...