128797 05
2017-08-11
21
5.4
Measuring Temperature
5.4.1 Using Thermocouples (TC)
The ASC-400 supports the following thermocouple types:
B, BP, C, E, J, K, L, N, R, S, T, U, XK
The characteristics of all the types are described in section 11 – Technical
Specifications.
The ASC-400 has 3 cold junction Modes to choose from:
CJ Auto
Automatic CJ compensation, CJ temperature inside the
connector is measured by a high accuracy RTD.
CJ Off
With CJ in OFF Mode, the calibrator will measure the difference
between the thermocouple at the junction and at its TC input
terminal (equivalent to 0ºC CJ).
CJ Manual
With CJ in Manual Mode it is possible to set the desired CJ
temperature to be used for temperature calculation.
Note…
CJ Off Mode should only be used when calibration is being done using an
external ice bath.
CJ Manual mode should be used when the cold junction temperature is known,
but different from 0 ºC.
To use the thermocouple to measure temperature, follow the principle of navigating
through the functions described in the guidelines in section 4.2 and proceed as follows
:
1. Attach the thermocouple leads to the TC mini plug, and insert the plug into the
input/output of the ASC-400 calibrator, as shown in Figure 4.
2. Enter edit mode.
3. Select “Thermocouple” from the Lower Mode list.
4. Select the desired Test Mode (In) from the Test Mode list.
5. Select TC function from the TC Type list.
6. Select CJ Mode from the list.
7. If CJ Manual is selected a CJ temperature value must be entered using either the
ARROW keys or the numeric keys.
8. Select the temperature unit from the Units list.
Содержание Jofra ASC-400
Страница 1: ...Reference manual Advanced Signal Calibrator Jofra ASC 400...
Страница 2: ...Reference Manual Advanced Signal Calibrator JOFRA ASC 400 Copyright 2014 AMETEK Denmark A S...
Страница 8: ...128797 05 2017 08 11 7 1 3 Dimensioning drawing 55 mm 96 mm 220 mm...
Страница 49: ...128797 05 2017 08 11 48 Figure 19 Calibrating an I P Device APM pressure module Pressure Device...
Страница 52: ...128797 05 2017 08 11 51 Figure 21 Calibrating a Pressure Transmitter Hand pump APM pressure module...