6.3
Sourcing Voltage
To source voltage, follow the principle of navigating through the functions described in
the guidelines in section 4.2 and proceed as follows
:
1. Select “Voltage” from the Lower Mode list.
2. Select “0 to 20 V” from the Range list.
3. Select “Out” from the Test Mode list.
4. Connect leads to the voltage source terminals, as shown in Figure 9.
5. Enter the desired voltage using either the ARROW keys or the numeric keys.
Figure 9
Connections for Sourcing
Voltage and Frequency
-
+
UUT