4. Raise the temperature slowly until the switch opens.
5. Lower the temperature slowly until the switch closes.
6. The upper display will now show :
the temperature when the switch closed
the temperature that the switch opened at
the dead band between the temperatures readings.
7. To perform another test clear the data by selecting “Reset” using the edit mode.