8.2.3
Reference sensor
This option enables you to view the calibration values currently
loaded in the DTI for the reference sensors as well as enter and
download values for new reference sensors. In addition to the
equipment already described in sections 8.1 and 8.2.1, you also
require the calibration certificate for the new reference sensor.
In DTI main menu, click “Sensor coefficients”.
A new dialogue with a table containing the coefficients
stored in the DTI for sensor 1 and sensor 2. Current sensor
coefficients are automatically uploaded.