7.3
Display and error messages
If the DTI detects an error while running, it will automatically appear
in the display. Following error messages can occur:
Display message:
Meaning/Action:
BATTERY LOW
Battery voltage low.
PRESS ANY KEY
Change batteries
(section 7.2).
No sensor connected to input
T1 ////.///°C
or measuring input outside
T2 ////.///°C
range.
Downloaded coefficients
CONSTANT ERRORS
rejected.
DEFAULT VALUES
Check calibration.
Default sensor coefficients
used are:
R0: +100.000000
A:+3.90802E-3
B : -5.80200E-7
C : -4.27350E-12
Press any key to continue