72 PET Plus User Manual
12.4 TESTING THE POLYMER
NoTE:
It is very important that the Die, the Barrel and the Follower are scrupulously clean
and have been assembled correctly after cleaning.
It is also important that the PET Plus is set to the “Pre-Test” screen ready for testing.
THIS PROCEDURE IS PART OF THE MAIN PROCEDURE SHOWN IN SECTION 7.7.
Press the key under PURGE to allow a low volume of nitrogen gas to flow through the barrel to
displace any air that may be present. Allow the gas to flow for a few seconds to ensure that the barrel
is full of nitrogen.
Press the key under GAS OFF to stop the flow of nitrogen gas.
Remove the bottom plug from the sample melt pot and position the melt pot onto the top of the barrel.
Remove the top plug from the sample melt pot and push the plug of polymer into the PET Plus barrel
using the charging tool.
Remove the sample melt pot from the top of the barrel.
Insert a clean follower into the top of the barrel.
Insert the charging tool into the top of the barrel.
Immediately press the green GO key to start the test. The “Charging” screen will be displayed.