PET Plus User Manual 69
11.5 USING THE POLYMER CHIP
TRANSFER VESSEL
Note that the production drying hopper MUST be
fitted with a chip outlet.
This outlet may be part of the flexible hose
assembly connected to the inlet of the extruder.
Turn the dispensing lever until the lever is
pointing upwards.
Ensure that the top and bottom caps are fitted.
Connect a low-pressure nitrogen supply to the
Schrader quick release inlet valve.
Set the external nitrogen pressure to approx 5psi
(0.35 Bar) and allow the nitrogen to flow through
the transfer vessel for about 10 minutes to make
sure the cylinder is completely dry.
Ensure that nitrogen gas flows out of the 5psi
pressure relief valve.
When ready to take a sample, disconnect the
nitrogen line from the Schrader quick release
inlet valve.
Do NOT remove the top or bottom caps.
Take the transfer vessel to the production area.
If a nitrogen supply is available at the sampling
site, either:
1
Reconnect the nitrogen to the Schrader
quick release inlet valve and allow the nitrogen
to flow for a minute.
2
Have a flexible hose that can be aimed
around the chip outlet.
Remove the top cap and load with the DRY
sample as quickly as possible.
The top chamber should be filled until it is be
-
tween half full and full of polymer.
The picture above shows a flexible hose
assembly connected to the inlet of the extruder.
IMMEDIATELY replace the top cap.
Take the transfer vessel containing the dry
sample back to the PET Plus.
The samples do not have to be tested immedi-
ately because they will be kept dry for several
hours if the top or bottom caps are not removed.
If the samples are to be stored for a time, it is
recommended that the low-pressure nitrogen
supply is reconnected to the Schrader quick
release inlet valve.