SECTION 2 – OPERATING INSTRUCTIONS 2-3
Cylinder Pressure Gauge Left/Right
Displays pressure inside the test cell.
Temperature Controller Left/Right
Controls temperature inside the test cell.
Heater Switch Left/Right
Turns the heater ON or OFF. Switch must be in the ON position during testing and should be
in the OFF position as a safety precaution at other times.
Main Power Switch
Turns power to the entire instrument ON or OFF. Switch must be in the ON position during
testing.
Top Transducer Connector
The coaxial connector attached to the top transducer must be connected to this connector
prior to the beginning of a test.
J Thermocouple Connector
The thermocouple that is attached to the top plug must be plugged into this connector prior to
the beginning of a test and any time that it is necessary for the temperature controller to
operate. If the thermocouple is not connected, the temperature will display
5.br
and will not
function.