Overview
|
1-13
The mathematics to map a specific range into another range are listed
below. This technique assumes that gain and offset are at their default
values, 1.0 and 0.0.
Actual ppm Values = Act_low and Act_high
Mapped Into Values = Map_low and Map_high
(Map_high
–
Map_low)
Gain
=
-------------------------------
(Act_high
–
Act_low)
Offset =
(Gain • Act_low) – Map_low
In the first example the calculations are:
Actual ppm Values = 50 ppm and 100 ppm
Mapped into Values = 0 ppm and 100 ppm
(Map_high
–
Map_low)
Gain
=
--------------------------------
(Act_high
–
Act_low)
Gain =
(100 – 0) / (100 – 50)
=
(100)
/
(50)
=
2.0
Offset =
(Gain • Act_low) – Map_low
=
(2.0 • 50) – 0
=
100.0
Since the goal is to get 50 to 100 ppm to be 0 to 100 percent on the 4 to
20-mA output, the above example could have been mapped into the 0 to
10-ppm range. The following shows the calculations for mapping 50 to
100 ppm into the 0 to 10-ppm range.
(Map_high
–
Map_low)
Gain
=
--------------------------------
(Act_high
–
Act_low)
Gain =
(10 – 0) / (100 – 50)
=
(100)
/
(50)
=
0.2
Offset =
(Gain • Act_low) – Map_low
=
(0.2 • 50) – 0
=
10.0
Содержание 303B
Страница 6: ...vi 303B Moisture Analyzer Manual 303B Moisture Monitor...
Страница 14: ...1 4 303B Moisture Analyzer Manual Figure 1 2 303B Flow Diagram 50 VDC...
Страница 24: ...1 14 303B Moisture Analyzer Manual This page intentionally left blank...
Страница 28: ...2 4 303B Moisture Analyzer Manual This page intentionally left blank...
Страница 36: ...3 8 303B Moisture Analyzer Manual Figure 3 2 Interconnection Diagram illustration...
Страница 37: ...Installation 3 9 Figure 3 3 Proper and Improper Grounding Methods...
Страница 39: ...Installation 3 11 Figure 3 4 Bubble O Meter PN 303030006...
Страница 40: ...3 12 303B Moisture Analyzer Manual This page intentionally left blank...
Страница 44: ...4 4 303B Moisture Analyzer Manual This page intentionally left blank...
Страница 46: ...5 2 303B Moisture Analyzer Manual Figure 5 1 Moisture Monitor Interior View...
Страница 60: ...6 4 303B Moisture Analyzer Manual Figure 6 1 Accessories...
Страница 62: ...A 2 303B Moisture Analyzer Manual This page intentionally left blank...
Страница 63: ...Appendix B B 1 Appendix B Figure B 1 Nomogram Conversion of PPm H2 0 to Relative Humidity...
Страница 64: ...B 2 303B Moisture Analyzer Manual This page intentionally leftr blank...
Страница 65: ...Appendix C C 1 Appendix C Figure C 1 Nomogram Determination of Dew Point with Pressure...
Страница 66: ...C 2 303B Moisture Analyzer Manual This page intentionally left blank...
Страница 68: ...D 2 303B Moisture Analyzer Manual This page intentionally left blank...
Страница 70: ...E 2 303B Moisture Analyzer Manual This page intentionally left blank...