Maintenance
|
5-3
Parts Replacement
Flow Meter Tubes (Figure 5-1)
1. Turn off power using front panel switch and disconnect the unit from
the power source.
2. Shut off sample flow at the source and purge the line. Shut off purge.
3. Remove the screws and lift off the cover.
4. Disconnect the “U” tube between the top fittings.
5. Remove the two screws and the plastic cover from the assembly.
6. With a hex wrench, loosen the plug over the glass tube in the assem-
bly top and remove the tube. Using a piece of tape on the glass makes
replacement easier.
7. Be sure the glass tube is centered on the top and bottom seals and the
grounding wire is around the tube before tightening the top plug.
Do not tighten the screws on the plastic cover or the cover might
break.
Flow Controller (Figure 5-1)
1. To gain access, perform steps 1, 2 and 3 above under Flow Meter Tubes
section.
2. Loosen the two tubing ferrule nuts at the flow controller and discon-
nect the lines.
3. Remove the mounting nuts and lift out the flow controller and brack-
et.
4. Remove the screw and remove the flow controller from the bracket.
Power Supply PC Board (Figure 5-1)
1. To gain access, perform steps 1, 2 and 3 above under Flow Meter Tubes
section.
2. Loosen the screws and remove the power-in wires from J1; remove
NOTE
Содержание 303B
Страница 6: ...vi 303B Moisture Analyzer Manual 303B Moisture Monitor...
Страница 14: ...1 4 303B Moisture Analyzer Manual Figure 1 2 303B Flow Diagram 50 VDC...
Страница 24: ...1 14 303B Moisture Analyzer Manual This page intentionally left blank...
Страница 28: ...2 4 303B Moisture Analyzer Manual This page intentionally left blank...
Страница 36: ...3 8 303B Moisture Analyzer Manual Figure 3 2 Interconnection Diagram illustration...
Страница 37: ...Installation 3 9 Figure 3 3 Proper and Improper Grounding Methods...
Страница 39: ...Installation 3 11 Figure 3 4 Bubble O Meter PN 303030006...
Страница 40: ...3 12 303B Moisture Analyzer Manual This page intentionally left blank...
Страница 44: ...4 4 303B Moisture Analyzer Manual This page intentionally left blank...
Страница 46: ...5 2 303B Moisture Analyzer Manual Figure 5 1 Moisture Monitor Interior View...
Страница 60: ...6 4 303B Moisture Analyzer Manual Figure 6 1 Accessories...
Страница 62: ...A 2 303B Moisture Analyzer Manual This page intentionally left blank...
Страница 63: ...Appendix B B 1 Appendix B Figure B 1 Nomogram Conversion of PPm H2 0 to Relative Humidity...
Страница 64: ...B 2 303B Moisture Analyzer Manual This page intentionally leftr blank...
Страница 65: ...Appendix C C 1 Appendix C Figure C 1 Nomogram Determination of Dew Point with Pressure...
Страница 66: ...C 2 303B Moisture Analyzer Manual This page intentionally left blank...
Страница 68: ...D 2 303B Moisture Analyzer Manual This page intentionally left blank...
Страница 70: ...E 2 303B Moisture Analyzer Manual This page intentionally left blank...