User Guide
5 - 8
Signal Processor
LMG MkII
5.2.3 RS485 examples
The RS485 communications is designed primarily for use by a computer
running a customised communications package.
To read the temperature from channel ‘A’ ...
Type at the terminal
[STX][addr]RATMP[ETX]
breakdown
STX Binary 0x02, start of text framing character
addr Binary address of processor to be interrogated
R
Read mode
A
Channel A
TMP Command to report the temperature
ETX Binary 0x03, end of text framing character
reply
[STX][addr]A1000C[ETX]
breakdown
STX Binary 0x02, start of text framing character
addr Binary address of replying processor
A
Channel ident of replying processor
1000C Temperature and units
ETX Binary 0x03, end of text framing character
To set the emissivity of channel ‘B’ to 0.5 ...
Type at the terminal
[STX][addr]SBEMS 50[ETX]
breakdown
STX Binary 0x02, start of text framing character
addr Binary address of processor to be interrogated
S
Set mode
B
Channel B
EMS Command to work on the emissivity value
50
New emissivity value
ETX Binary 0x03, end of text framing character
reply
Will reply with “[channel letter]*” if command fails
Содержание LMG MkII
Страница 6: ...INTRODUCTION 1 ...
Страница 10: ...INSTALLING THE PROCESSOR 2 ...
Страница 14: ...User Guide 2 4 Signal Processor LMG MkII Fig 2 4 LMG MkII Electrical System Overview ...
Страница 22: ...3 USING THE PROCESSOR ...
Страница 44: ...TIME FUNCTION PROCESSING 4 ...
Страница 50: ...SERIAL COMMUNICATIONS 5 ...
Страница 62: ...Signal Processor LMG MkII Blank ...
Страница 63: ...EXTERNAL I O MATHS FUNCTIONS 6 ...
Страница 64: ...Signal Processor LMG MkII Screws A Fig 6 1 Location of screws on back panel Screws A ...
Страница 88: ...MAINTENANCE 7 ...
Страница 92: ...User Guide 7 4 Signal Processor LMG MkII Fig 7 3 USB Export Data option ...
Страница 94: ...UPGRADING FROM LMG MkI TO LMG MkII 8 ...