![Ametek Land LMG MkII Скачать руководство пользователя страница 51](http://html1.mh-extra.com/html/ametek-land/lmg-mkii/lmg-mkii_user-manual_2931447051.webp)
5 - 1
User Guide
LMG MkII
Signal Processor
5
Serial Communications
The Landmark Graphic MkII RS485 serial port allows communication to set
read parameters to the processor.
The port must be set up to match the communication parameters of the
sending device. The following parameters are setable (See Fig. 5-1).
Access to the Serial settings screen is via the main menu (See Fig. 3-3).
Unit address:
0 to 15
Baud rate:
300, 600, 1200, 2400, 4800, 9600, 57600 and 115200
Data bits:
8 or 7 bits
Parity:
None, Odd or even
Stop bits:
1 or 2
Table 5-1 Serial Communications specification
Fig. 5-1 Serial Port Settings screen
The available commands are given in Section 5.1.
The detailed format of the commands is given in Section 5.2.
Specific timing requirements are given Section 5.3.
Содержание LMG MkII
Страница 6: ...INTRODUCTION 1 ...
Страница 10: ...INSTALLING THE PROCESSOR 2 ...
Страница 14: ...User Guide 2 4 Signal Processor LMG MkII Fig 2 4 LMG MkII Electrical System Overview ...
Страница 22: ...3 USING THE PROCESSOR ...
Страница 44: ...TIME FUNCTION PROCESSING 4 ...
Страница 50: ...SERIAL COMMUNICATIONS 5 ...
Страница 62: ...Signal Processor LMG MkII Blank ...
Страница 63: ...EXTERNAL I O MATHS FUNCTIONS 6 ...
Страница 64: ...Signal Processor LMG MkII Screws A Fig 6 1 Location of screws on back panel Screws A ...
Страница 88: ...MAINTENANCE 7 ...
Страница 92: ...User Guide 7 4 Signal Processor LMG MkII Fig 7 3 USB Export Data option ...
Страница 94: ...UPGRADING FROM LMG MkI TO LMG MkII 8 ...