Plate Mounted Blower Unit
User Guide
Page 18
11 Maintenance
The Plate-mounted blower unit has been designed for a minimal amount of
routine maintenance. It is important to follow the recommended schedule. This
schedule should be run alongside that recommended for the instrument itself.
Maintenance Schedule
Item
Part
Maintenance
Procedure
Air Hose
318.239
90 Days
Examine hoses for holes or leaks.
Test all hose clamps for tightness
Pre-Filter
(where fitted)
317.560
90 Days
Dust level should be below the line.
Remove, empty, wipe out, and replace.
Safety Filter
(where fitted)
317.604
90 Days
Safety filter will require changing when
used. Inspect as specified. Always re-
place old for new, do not try to clean.
Filter Element
317.561
90
Annually
Wear eye protection, tap gently to re-
move dust. Clean carefully with a blow
gun.
Examine for holes or rips. Re-fit.
Replace with new filter.
Blower
90 Days
Check for bearing noise. (Bearing failure,
and subsequent blower failure, can be
detected early.) Check blower air flow
to be sure that the blower is performing
normally. (Dirty air filters may reduce air
flow and cause undue stress