ThePoint Series
20
2.10.4 Accessing the Calibration Modes
1.
On the top side of ThePoint, temporarily remove the shunt
from the “Time Delay Selection Jumper” (see Fig. 2) and
place it on pins 1 & 2 of the 3-pin connector. Pin 1 is closet
to the LEDS. The green LED will go out and the red LED
will begin to flash. The number of flashes indicates which
mode the unit is in (1 through 8).
2.
To switch modes, press and hold the ReCal button next to
the 3-pin connector. The unit will cycle through the modes.
First it will flash the current mode setting, then progress
through all of the settings.
For Example:
The red LED will flash once indicating mode 1. Then it will
flash twice-indicating mode 2. Then mode 3, etc.
Release the button when it reaches the desired mode. The
red LED will now flash indicating which mode the unit is
in.
3.
Remove the shunt from pins 1 & 2 on the 3-pin connector
and replace the shunt on the “Time Delay Selection Jumper”.
The unit will remain in the selected mode. Write the new
mode # on the inside of the lid label for future reference
4.
After setting the mode follow procedure in section 2.8.2
for mode 2 & 6. For modes 7 and 8, follow the appropriate
manual calibration procedure as described in section 2.10.3.
Electronic Unit Adjustments
Do not push
the Recal button
without first ensuring
that the material
being measured is
below the sensing
element.
Failsafe Jumper
Up = High Level
Down = Low Level
Time Delay Jumper
Up = Forward Acting
Down = Reverse Acting
ReCal Button
3 Pin Connector
Содержание ThePoint Series
Страница 4: ...Contents...
Страница 6: ...Section 1...
Страница 28: ...ThePoint Series 22 Section 3...
Страница 36: ...Section 4...
Страница 39: ...Control Drawings 33 Section 5 Control Drawings 5 1 FM FMc Control Drawings...
Страница 40: ...ThePoint Series 34 5 1 FM FMc Control Drawings Continued...
Страница 41: ...Control Drawings 35 5 1 FM FMc Control Drawings Continued...
Страница 42: ...ThePoint Series 36 5 1 FM FMc Control Drawings Continued...
Страница 43: ...Control Drawings 37 5 1 FM FMc Control Drawings Continued...
Страница 44: ...ThePoint Series 38 5 1 FM FMc Control Drawings Continued...
Страница 45: ...Control Drawings 39 5 1 FM FMc Control Drawings Continued...
Страница 46: ...ThePoint Series 40 5 1 FM FMc Control Drawings Continued...
Страница 47: ...Control Drawings 41 5 1 FM FMc Control Drawings Continued...
Страница 48: ...ThePoint Series 42 5 1 FM FMc Control Drawings Continued...
Страница 49: ...Control Drawings 43 5 1 FM FMc Control Drawings Continued...
Страница 50: ...ThePoint Series 44 5 2 ATEX Control Drawings...
Страница 51: ...Control Drawings 45 5 2 ATEX Control Drawings Continued...
Страница 52: ...ThePoint Series 46 5 2 ATEX Control Drawings Continued...
Страница 55: ...Control Drawings 49 5 2 ATEX Control Drawings Continued...
Страница 56: ...ThePoint Series 50 5 3 Heavy Duty Spark Protection...
Страница 57: ...Control Drawings 51 5 3 Heavy Duty Spark Protection Continued...
Страница 58: ...ThePoint Series 52 5 4 Adding a Padded Capacitor...
Страница 59: ...Control Drawings 53 5 4 Adding a Padded Capacitor Continued...
Страница 60: ...ThePoint Series 54 5 4 Adding a Padded Capacitor Continued...
Страница 61: ...Control Drawings 55 5 5 Dual Seal Assembly for 700 Series Sensing Elements...
Страница 62: ...ThePoint Series 56 5 5 Dual Seal Assembly Continued...
Страница 63: ...Control Drawings 57 5 5 Dual Seal Assembly Continued...
Страница 64: ...Appendix A...
Страница 69: ......