20
5.4
STANDARD SETPOINTS
FACTORY AS
SETPOINT
DESCRIPTION
RANGE
SETTING
SHIPPED
TDS/Cond Limit
When this value is met or exceeded,
0-999
Based on
45
the alarm lamp will light and high
µS or PPM* water
TDS/Cond will show on the display.
analysis.
To disable, set to 0.
TDS/Cond Delay
When the limit Setpoint is exceeded,
0-999
10
180
no alarm will be given until this time
seconds
has expired.
RO Start Delay
The amount of time between the
0-99
10
80
inlet valve opening and the RO pump. Seconds
start.
Press Fault Delay The time a pressure fault must be
0-99
10
30
active before a pressure fault shut
Seconds
down occurs.
Auto Reset
When a pressure fault shut down is
0-99
0
60
active,the system will attempt to restart minutes
after this delay. If set to 0, system
must be manually reset.
Alarm Silence
If the audible alarm is silenced, after
0-99
3
0
this delay, the alarm will resound. If
minutes
set to 0, the alarm will remain silenced.
TF Restart Delay When a tank full condition clears,
0-99
5
5min
the system will restart after this delay. sec/min
TF Restart
Selects whether the tank full restart
0-1
0
0
delay is in seconds or minutes.
0=seconds, 1=minutes.
TFO Time
The amount of time that a tank full
0-9
3
5min
override lasts.
Minutes
Tank Lo Restart
Not Used
15min
Flush Type
Selects the type of flush. Set to 0
0-8
0
0
to disable.
Flush Time
The length of time a membrane flush
0-99
0
0
cycle will last when flush is active.