KMM1002P
43
Disconnect the fan connector, plug system power in, turn the
unit on, press top cover interlock switch when applicable,
and check for 24 VDC in the harness side of the connector.
If 24 VDC is present, then change the fan. If no voltage is
present, replace the power supply.
Timer Not Running
(Power is on, fans are running and front panel is on.)
Verify that the footswitch is plugged in.
Unplug the footswitch. With the footswitch depressed, check
the male plug contacts for continuity. If there is no continuity
when the footswitch is depressed, replace the footswitch.
Содержание Lesco SUPERSPOT MK III
Страница 4: ...KMM1002P 2...
Страница 33: ...KMM1002P 31...
Страница 34: ...KMM1002P 32...
Страница 35: ...KMM1002P 33...
Страница 40: ...KMM1002P 38 7 2 5 Shutter Activate Input Wiring 1 1 2 2 FOOT SWITCH SOLID STATE 250mA INDUCTIVE...
Страница 46: ...KMM1002P 44 Appendix A Agency Approvals...
Страница 47: ...KMM1002P 45...
Страница 48: ...KMM1002P 46...
Страница 49: ...KMM1002P 47...
Страница 50: ...KMM1002P 48 Appendix B UV Protective Eye Glasses...