KMM1002P 

 

       

28

      

quote  a  variety  of  custom  guides  for  special  purposes.    These  include  special  models 
optimized for high transmission in the UV “B” and UV “C” range (230 to 390 nm).  Contact 
AUV/LESCO factory for price and delivery of special products. 

AUV/LESCO’s precision-built LF lightguides are high quality optical instruments, but their 
performance  will  naturally  degrade  with  time.    If  cared  for  properly  and  under  normal 
operating  conditions,  the  user  can  expect  anywhere  from  perhaps  1  to  2  years  of  use 
from  a  LF  lightguide.    However,  normal  degradation  will  be  greatly  accelerated  with 
aggressive  radiation  and/or  extreme  heat  levels.    For  a  given  dosage,  it  is  well 
established that high intensity for short exposure time periods is the preferred method of 
effecting  maximum  cure  strength.  This  is  true  for  the  great  majority  of  applications  and 
will not normally damage the guides.  

However, care must be taken to minimize very high intensity levels for sustained periods 
of exposure time.  Also, the special IR reduction filter in AUV/LESCO SUPERSPOT cure 
systems must not be removed.  If either of these conditions occurs, the LF lightguide will 
suffer  irreparable  damage  or  have  its  service  life  reduced  dramatically.    Special  duty 
AUV/LESCO  FO  lightguides  are  available  for  extreme  intensity/dosage  applications 
(contact factory at 1-800-615-3726).

  

Never  bend  a  light  guide  radically  (i.e.  -  over  90

°

  over  a  12”  radius  for  a  LF).    Be 

extremely  careful  not  to  scratch,  mar or accidentally apply curable material to the guide 
end  tips.  Clean  the  quartz  end  tips  with  alcohol  on  a  regular  basis  and  check 
transmission  levels  at  least  every  100  operating  hours  (more  often  if  end  tips  are  not 
fixtured).  

 

6.10. Fiber-optic 

AUV/LESCO fiber optic (FO) lightguides are used for applications where long run lengths, 
tight  end  clearances,  or  multiple  outputs  are  required.    FO  lightguides  are  available  in 
almost  endless  configurations  for  specialized  applications.    Popular  standard  products 
are  Bi-,  tri-  and  quadfurcated  guides,  although  AUV/LESCO  FO  lightguides  can  be 
ordered  with  10  or  more  output  ends  arranged  in  circular,  line  or  any  random 
configurations. 

Your  Fiber  Optic  Light  Guide  (FOLG)  is  designed  to  provide  a  long  life  of  dependable 
and consistent UV energy transmission. In order to sustain the high levels of throughput 
and extend its useful life, the FOLG must be properly maintained.  
 

Following  are  some  general  maintenance  guidelines  for  your  reference.  If  you  need  any 
additional help or have further questions, please contact AUV/LESCO with your request.  
 

1.  Inspect the FOLG regularly. A visual inspection of its ends will detect most problems 

associated  with  build  up  of  residues  or  signs  of  deterioration.  To  inspect  it,  use  a 
bright  incandescent  lamp.  It  is  better  to  cover  the  opposite  end  to  prevent  back 
illumination. The surface should look black and very shiny. If it looks dull or not glossy, 
there is residue on it and it should be cleaned.  

2.  Never use a sharp object to clean the fiber surfaces. Use only high quality solvents. 

We recommend using reagent grade acetone. Do not use regular acetone as it may 
leave residues.  

Содержание Lesco SUPERSPOT MK III

Страница 1: ...STEM Model VSM3003 TECHNICAL REFERENCE MANUAL AUV LESCO A Division of American Ultraviolet Company A WWW LESCOUV COM 23555 Telo Avenue Torrance CA 90505 10 6 10 This product conforms to CE Standards S...

Страница 2: ...Any repairs made after the original equipment warranty period has expired are warranted for ninety 90 days from the date the repaired equipment is shipped back to the purchaser After all warranties ha...

Страница 3: ...e returned at the purchaser s expense 3 All returns must be correctly packaged and shipped via an appropriate courier AUV West LESCO is not liable for any damage or charges incurred as a result of imp...

Страница 4: ...KMM1002P 2...

Страница 5: ...ly 11 3 2 5 Shutter Assembly 12 3 2 6 Timer Manual Selector Switch 12 3 2 7 Footswitch 12 3 2 8 Power Cord 12 3 3 BACK PANEL 13 3 3 1 Cooling Fans 13 3 3 2 Power Control Switch 14 3 3 3 IR Filter 14 3...

Страница 6: ...CTOR ASSEMBLY 27 6 6 SHUTTER 27 6 7 IR FILTER 27 6 8 LIGHTGUIDES 27 6 9 LIQUID FILLED 27 6 10 FIBER OPTIC 28 6 10 1 All Lightguides 29 7 0 DRAWINGS SCHEMATICS 30 7 2 INPUT OUTPUT CONNECTOR WIRING 34 7...

Страница 7: ...ght source whose output is collected and focused into one end of a flexible light guide The shutter is located between the light source and the lightguide entrance point The shutter is activated by ei...

Страница 8: ...lenses or remain behind UV blocking light shielding Personnel should also be prevented from allowing any skin surface from coming within close proximity of the high intensity light output s Additional...

Страница 9: ...or more efficient use of counter top space The optical assembly is located in the upper compartment The top cover is secured by loosening four quarter turn captive Phillips head fasteners The power su...

Страница 10: ...ains the optical transmission assembly bulb reflector optical feedback sensor and electrical lamp connections The function of each component in the lamp housing is described below Reflector Lightguide...

Страница 11: ...it impossible to assign a conclusive value to bulb life under intermittent operation As a general rule it is recommended that if the unit is to be used again within the next 4 hours it should be left...

Страница 12: ...of energy from the bulb 3 1 5 Bulb Reflector Module The Bulb Reflector Module is a detachable assembly consisting of the bulb metal holder and an elliptical reflector The bulb is optically aligned wi...

Страница 13: ...t also supports the shutter assembly optical feedback sensor lightguide safety switch and infrared filter The lightguide transmits UV energy most efficiently when it is straight Bends in the lightguid...

Страница 14: ...he MK III 3 2 6 Timer Manual Selector Switch The Timer Manual toggle switch is located next to the bottom cooling fan on the recessed panel at the rear of the unit In the Timer position pressing the f...

Страница 15: ...ots 3 3 1 Cooling Fans The fan system is used to maintain the optimum temperature of the Lamp while cooling the optics and electronics The bottom fan is used for intake and the upper fan is used for e...

Страница 16: ...ns The filter function is to screen out the unwanted radiation from the focused light provided by the reflector thus protecting the lightguide as well as the work surface from excessive heat 3 3 4 Opt...

Страница 17: ...ptional Footswitch with 6 foot lead and phone jack UV Protection Goggles SUPERSPOT MK III Manual The MKIII must never be placed with the front panel facing downwards This places the cathode of the lam...

Страница 18: ...erating and the shutter is open a concentrated beam of UV light is emitted from the light exit port UV light is hazardous to the eyes and skin Never open the shutter without wearing protective goggles...

Страница 19: ...he chassis This will allow access into the bulb reflector assembly 3 Remove the packing material from the bulb and the nut from the bulb base Then insert the bulb into the base of the reflector Be car...

Страница 20: ...ok into the emitting end of the lightguide Use proper shielding at all times Safety glasses are provided with each system Avoid exposing skin to UV light Shielded UV enclosures and safety glasses are...

Страница 21: ...er line the MKIII will be connected to Remove the bottom cover by undoing the six button head screws The voltage selector connector is located on the motherboard of the power supply near the front of...

Страница 22: ...re 1 Detail A A To change from 120VAC to 240VAC input move the connector shown in Detail A A 1 pin position forward as shown above Figure 1A Voltage Setting See Detail A A Voltage Selection Connector...

Страница 23: ...essure from the footswitch if in Manual mode or wait for timed shutter open period to expire 4 4 3 Proper Light Exposure UV light is most effective when it is perpendicular to the work surface the eff...

Страница 24: ...t will be emitted from light exit port The shutter will close automatically when the preset time has elapsed 4 5 3 Proper Light Exposure UV light is most effective when it is perpendicular to the work...

Страница 25: ...be set by adjusting the potentiometer located in the rear of the unit while measuring the output of the lightguide with a calibrated NIST traceable radiometer Turning the potentiometer counter clockwi...

Страница 26: ...ossible from the upper output limit This would allow more reserve to compensate for bulb output decay 5 2 Optical Feedback Problems 1 Setting the reference voltage too low outside the operating range...

Страница 27: ...cohol Let unit air dry before applying power 6 2 1 Air Filter Cleaning The MK III has two fans located at the rear panel each of the fans is equipped with a removable air filter To remove the air filt...

Страница 28: ...A in first line Jumpered in second line Spare fuse 4A For 220vac Fuse 2A in first line Remove the jumper Install a 2A fuse in second line Correct fuse placement in holder No fuse Correct Incorrect Con...

Страница 29: ...it should be replaced to prevent failures or mis operation every 10 million operations 6 7 IR Filter The IR filter is a specially coated high quality quartz plate The filter is a highly critical compo...

Страница 30: ...r or accidentally apply curable material to the guide end tips Clean the quartz end tips with alcohol on a regular basis and check transmission levels at least every 100 operating hours more often if...

Страница 31: ...se serious damage to the fiber Following these general guidelines and proper preventive maintenance will give you the expected long life of your FOLG Please make sure you keep it clean and working pro...

Страница 32: ...e MKIII system and accessories 7 1 Main Schematic MKIII with VUM1022 HP Interface MKIII with VUM1022 Interface 7 2 Input Output Connector Wiring I O Interface Option VUM1022 HP I O Interface Option VU...

Страница 33: ...KMM1002P 31...

Страница 34: ...KMM1002P 32...

Страница 35: ...KMM1002P 33...

Страница 36: ...PUT PULSING 1Hz WHEN ALARM IS ON INPUT COMMON INPUT CONTINUITY BETWEEN PINS 1 2 WILL ACTIVATE SHUTTER OUTPUT PIN 3 18VDC TO PIN 6 WHEN POWER ALERT CONDITION EXISTS 24VDC SUPPLY 25mA MAX OUTPUT 24 VDC...

Страница 37: ...G 1Hz WHEN ALARM IS ON INPUT CONTINUITY BETWEEN PINS 1 2 WILL ACTIVATE SHUTTER INPUT COMMON OUTPUT PIN 3 TO PIN 6 WHEN POWER ALERT CONDITION EXISTS OUTPUT PINS 4 TO PIN 6 WHEN LAMP ON 24VDC SUPPLY 25m...

Страница 38: ...T CONTINUITY BETWEEN PINS 1 6 WILL ACTIVATE SHUTTER SHUTTER SAFETY INTERLOCK 250mA INDUCTIVE LOAD OUTPUT PIN 3 TO PIN 6 WHEN POWER ALERT CONDITION EXISTS SHUTTER SAFETY INTERLOCK 250mA INDUCTIVE LOAD...

Страница 39: ...TION FUNCTION 2 1 3 5 7 6 4 NOTE LIMIT CURRENT THROUGH OUTPUTS 3 4 TO 30mA MAX OUTPUT PULSING 1Hz WHEN ALARM IS ON OUTPUT PIN 3 TO PIN 6 WHEN POWER ALERT CONDITION EXISTS OUTPUT COMMON OUTPUT PINS 4 T...

Страница 40: ...KMM1002P 38 7 2 5 Shutter Activate Input Wiring 1 1 2 2 FOOT SWITCH SOLID STATE 250mA INDUCTIVE...

Страница 41: ...er AIB1001 8 2 Optional Liquid filled Lightguides 5mm x 1M OLB1003 5mm x 1 5M OLB1004 5mm x 3M OLB1009 Dual 2 x 3mm x 1M OLB1001 Dual 2 x 3mm x 1 5M OLB1002 Trifurcated 3 x 3mm x 1 5M OLB1012 8mm x 1M...

Страница 42: ...voltage configuration 120 220 VAC If unit is configured for 120VAC and 220VAC power is applied the unit will blow fuses Power supply may be damaged Check AC voltage jumper and apply correct voltage R...

Страница 43: ...itch is plugged in if applicable and is in proper operation following procedures in section 9 5 above Turn power OFF Remove bottom cover Disconnect solenoid connector from main harness Verify shutter...

Страница 44: ...or AUV LESCO website WARNING Bulbs are extremely hot when first shut off and can cause severe burns Disconnect and remove the lamp inspect the bulb visually Note Do not touch quartz surfaces particul...

Страница 45: ...tor If 24 VDC is present then change the fan If no voltage is present replace the power supply Timer Not Running Power is on fans are running and front panel is on Verify that the footswitch is plugge...

Страница 46: ...KMM1002P 44 Appendix A Agency Approvals...

Страница 47: ...KMM1002P 45...

Страница 48: ...KMM1002P 46...

Страница 49: ...KMM1002P 47...

Страница 50: ...KMM1002P 48 Appendix B UV Protective Eye Glasses...

Страница 51: ...tput graphs of our various filters on the UV Spot Cure series Filter OFB1004 280nm 480nm 0 1000 2000 3000 4000 200 300 400 500 600 700 800 Intensity counts Wavelength nm Filter OFB1008 280nm 400nm 0 1...

Страница 52: ...lter OFB1013 275nm 500nm 0 1000 2000 3000 4000 200 300 400 500 600 700 800 Intensity counts Wavelength nm Filter OFB1014 280nm 320nm 0 1000 2000 3000 4000 200 300 400 500 600 700 800 Intensity counts...

Страница 53: ...KMM1002P 51 Filter OFB1019 400nm 500nm Filter OFB1021 320nm 420nm 0 1000 2000 3000 4000 200 300 400 500 600 700 800 Intensity counts Wavelength nm...

Страница 54: ...FOR SPARE PARTS SALES OR SERVICE CALL AUV LESCO ADivisionofAmericanUltravioletCompany 23555 TELO AVENUE TORRANCE CA 90505 USA PHONE 310 784 2930 800 615 3726 FAX 310 784 2929 800 905 3726 E mail Sales...

Отзывы: