American Diagnostic Corporation Satin Скачать руководство пользователя страница 7

7

Battery Replacement: 

(4067F Stubby Handle)

1)  Unscrew bottom cap of handle and remove battery insert.

2)  Replace with appropriate size batteries. Insert 1 battery positive

end first and the other negative end first into battery insert, paying

attention to the polarity markings. Slide battery insert back into han-

dle following the direction of the “Insert This Way Only” arrows.

Screw bottom cap on and tighten. (Figure 3)

NOTE:

Alkaline batteries should be used with your handle for maxi-

mum performance and are recommended for replacement. Ordinary

carbon zinc batteries may also be used.

Lamp Replacement: 

(Standard Models)

For correct illumination, alignment, and assurance of watertight seal during

immersion/autoclave, we recommend using only ADC #4500-1 and #4501-1 replace-

ment lamps for standard blades.

1)  To remove lamp on standard blades, grip lamp collar and rotate lamp counterclock-

wise until free.

2)  Replace the lamp with a new ADC lamp.

3)  Verify that lamp is sufficiently tightened before use.

CAUTION

: Be sure lamp’s glass surface is clean and free from fingerprints after assem-

bly. If necessary, the glass may be cleaned with a soft cloth or cotton ball moistened in alcohol.

Lamp Replacement: 

(Fiberoptic Models)

For correct illumination and alignment we recommend using only

ADC replacement lamp model #5111N-4 (Figure 4).

1)  Prior to removal, allow lamp time to cool.

2)   Unscrew the laryngoscope head from the barrel of the handle in a

counterclock-wise direction. 

3)  Pull lamp out of the laryngoscope head. The lamp is friction fit and does not need to

be twisted.

4)  Insert replacement lamp by firmly pressing into head until it is completely seated.

5)  Screw the laryngoscope head onto the barrel of the handle. Tighten by hand only.

Main

Handle

Battery

Insert

AA Batter

y

AA Batter

y

Insert This W

ay Only

Figure 3

Main

Handle

Battery

Insert

A

A

 B

a

tt

e

ry

A

A

 B

a

tte

ry

Figure 4

In

s

e

rt

 T

h

is

 W

a

y

 O

n

ly

Содержание Satin

Страница 1: ...ADC SatinTM Standard Fiber Optic and Disposable Laryngoscope Blades and Handles Directions for Use...

Страница 2: ...ications The following are only relative contraindication to tracheal intubation 1 Severe airway trauma or obstruction that does not permit safe passage of an endotra cheal tube Emergency cricothyroto...

Страница 3: ...ation of all Laryngoscope Models To Activate Attach selected blade onto handle by hooking blade onto handle rod as shown in Figure 1 Pull blade up ward to open position to automatically activate light...

Страница 4: ...aning and replacement WARNING Ultrasonic cleaning is not recommended Disinfection Sterilization Warnings for All Laryngoscope Blades CAUTION ADC laryngoscope blades do not come sterilized The Sterrad...

Страница 5: ...be wiped with a cloth dampened with 70 isopropyl alcohol DO NOT ALLOW SOLUTION TO ENTER INTO THE CARTRIDGE The lamp cartridge and bulb CANNOT be SOAKED OR AUTO CLAVED High Level Disinfection for Stan...

Страница 6: ...OLUTION TO ENTER INTO THE CARTRIDGE The lamp cartridge and bulb CANNOT be SOAKED OR AUTOCLAVED Steris Sterilization Processes for Disposable and Fiberoptic Laryngoscopes This product has been validate...

Страница 7: ...mp collar and rotate lamp counterclock wise until free 2 Replace the lamp with a new ADC lamp 3 Verify that lamp is sufficiently tightened before use CAUTION Be sure lamp s glass surface is clean and...

Страница 8: ...y reversing the above steps 4 Tighten the locking screw ensuring that the fiberoptic bundle is securely attached to the laryngoscope blade Figure 5 Test Procedures Laryngoscope blades and handles shou...

Страница 9: ...ELS DESCRIPTIONS Laryngoscope Item Mac 0 Mac 1 Mac 2 Mac 3 Mac 4 Miller 00 Miller 0 Miller 1 Miller 2 Miller 3 Miller 4 Wisconsin 0 Wisconsin 1 Wisconsin 1 5 Wisconsin 2 Wisconsin 3 Wisconsin 4 C Cell...

Страница 10: ...ll Handle Stubby Handle Fiberoptic Blade 4070F 4071F 4072F 4073F 4074F NA 4080F 4081F 4082F 4083F 4084F NA NA NA 4NA NA NA 4065F 4066F 4067F Fiberoptic Replacement Lamp NA NA NA NA NA NA NA NA NA NA N...

Страница 11: ...84D 20 NA NA NA NA NA NA NA 4065D 20 NA SPU MODELS DESCRIPTIONS REPLACEMENT PARTS CHART Laryngoscope Item Mac 0 Mac 1 Mac 2 Mac 3 Mac 4 Miller 00 Miller 0 Miller 1 Miller 2 Miller 3 Miller 4 Wisconsin...

Страница 12: ...ental special or consequential damages so this limitation may not apply to you To Obtain Warranty Service Send item s postage paid to ADC Attn Repair Dept 55 Commerce Dr Hauppauge NY 11788 Please incl...

Страница 13: ...it 6 PO14 1TH United Kingdom Inspected in the U S A Reusable models Made in Pakistan Disposable models Made in Taiwan China tel 631 273 9600 toll free 1 800 232 2670 fax 631 273 9659 www adctoday com...

Отзывы: