Basic Cleaning Procedure for Standard and Fiberoptic Blades:

Prepare a Klenzyme

®

, Manu-klenz

, or equivalent cleaning solution as per manufac-

turer’s instructions and soak for ten minutes. Rinse devices with lukewarm running tap

water for one minute to remove any residual detergent after cleaning. Dry with a clean,

lint-free cloth.

                                                   

Basic Cleaning Procedures for Standard and Fiberoptic Handles:

Prepare a Klenzyme

®

, Manu-klenz

, or equivalent cleaning solution as per manufacturer’s in-

structions. Remove both end caps, lamp cartridge and bulb, and batteries from device, and

rinse with lukewarm tap water until all visible contamination has been removed. Soak handle

and end caps in the prepared solution for a minimum of ten minutes, ensuring that all compo-

nents are fully submerged. After the soak, the components should be scrubbed with a soft

bristled brush paying particular attention to hard to reach areas. Rinse with cool running tap

water to remove all residual detergent. Dry with a clean, lint free cloth. Note: Always allow

handles to completely dry before disinfection or use. Lamp cartridge and bulb can be wiped

with a cloth dampened with 70% isopropyl alcohol. DO NOT ALLOW SOLUTION TO ENTER

INTO THE CARTRIDGE. The lamp cartridge and bulb CANNOT be SOAKED OR AUTO-

CLAVED.

High-Level Disinfection for Standard and Fiberoptic Blades:

To achieve high-level disinfection, a 2.4% glutaraldehyde solution (Cidex

®

or equivalent)

must be prepared at 77°F (25°C) to the manufacturer’s instructions. The blades should be

fully immersed in the disinfectant for a minimum of 45 minutes. No air bubbles should be

visible on the immersed blades during this soak time. The blades should then be removed

from the disinfectant and immersed in purified water to rinse off all disinfectant. Immersion

should be for a minimum of one minute and the process should be repeated two additional

times using fresh purified water each time. Once rinsing is complete, the blades should be

thoroughly dried using a sterile, lint-free cloth. Note: Always allow blades to completely dry

before use.

High-Level Disinfection for Disposable Fiberoptic Blades:

Disposable fiberoptic laryngoscope blades are supplied non-sterile. They are designed for

single patient use only and are not disinfected after use. Gas sterilization by Ethylene oxide

following these guidelines. Temperature 55°C (±5°C) EO pressure - 180 mmHg (± 20

mmHg) Chamber pressure 210mmHg (±20mmHg).

Содержание Satin

Страница 1: ...ADC SatinTM Standard Fiber Optic and Disposable Laryngoscope Blades and Handles Directions for Use...

Страница 2: ...ications The following are only relative contraindication to tracheal intubation 1 Severe airway trauma or obstruction that does not permit safe passage of an endotra cheal tube Emergency cricothyroto...

Страница 3: ...ation of all Laryngoscope Models To Activate Attach selected blade onto handle by hooking blade onto handle rod as shown in Figure 1 Pull blade up ward to open position to automatically activate light...

Страница 4: ...aning and replacement WARNING Ultrasonic cleaning is not recommended Disinfection Sterilization Warnings for All Laryngoscope Blades CAUTION ADC laryngoscope blades do not come sterilized The Sterrad...

Страница 5: ...be wiped with a cloth dampened with 70 isopropyl alcohol DO NOT ALLOW SOLUTION TO ENTER INTO THE CARTRIDGE The lamp cartridge and bulb CANNOT be SOAKED OR AUTO CLAVED High Level Disinfection for Stan...

Страница 6: ...OLUTION TO ENTER INTO THE CARTRIDGE The lamp cartridge and bulb CANNOT be SOAKED OR AUTOCLAVED Steris Sterilization Processes for Disposable and Fiberoptic Laryngoscopes This product has been validate...

Страница 7: ...mp collar and rotate lamp counterclock wise until free 2 Replace the lamp with a new ADC lamp 3 Verify that lamp is sufficiently tightened before use CAUTION Be sure lamp s glass surface is clean and...

Страница 8: ...y reversing the above steps 4 Tighten the locking screw ensuring that the fiberoptic bundle is securely attached to the laryngoscope blade Figure 5 Test Procedures Laryngoscope blades and handles shou...

Страница 9: ...ELS DESCRIPTIONS Laryngoscope Item Mac 0 Mac 1 Mac 2 Mac 3 Mac 4 Miller 00 Miller 0 Miller 1 Miller 2 Miller 3 Miller 4 Wisconsin 0 Wisconsin 1 Wisconsin 1 5 Wisconsin 2 Wisconsin 3 Wisconsin 4 C Cell...

Страница 10: ...ll Handle Stubby Handle Fiberoptic Blade 4070F 4071F 4072F 4073F 4074F NA 4080F 4081F 4082F 4083F 4084F NA NA NA 4NA NA NA 4065F 4066F 4067F Fiberoptic Replacement Lamp NA NA NA NA NA NA NA NA NA NA N...

Страница 11: ...84D 20 NA NA NA NA NA NA NA 4065D 20 NA SPU MODELS DESCRIPTIONS REPLACEMENT PARTS CHART Laryngoscope Item Mac 0 Mac 1 Mac 2 Mac 3 Mac 4 Miller 00 Miller 0 Miller 1 Miller 2 Miller 3 Miller 4 Wisconsin...

Страница 12: ...ental special or consequential damages so this limitation may not apply to you To Obtain Warranty Service Send item s postage paid to ADC Attn Repair Dept 55 Commerce Dr Hauppauge NY 11788 Please incl...

Страница 13: ...it 6 PO14 1TH United Kingdom Inspected in the U S A Reusable models Made in Pakistan Disposable models Made in Taiwan China tel 631 273 9600 toll free 1 800 232 2670 fax 631 273 9659 www adctoday com...

Отзывы: