High-Level Disinfection for Standard and Fiberoptic Handles:

To achieve high-level disinfection, a 2.4% glutaraldehyde solution (Cidex

®

or equivalent)

must be prepared at 77°F (25°C) to the manufacturer’s instructions. The end caps of the

handles, lamp cartridge and bulb, and batteries must be removed. Both the end caps and

handles should be fully immersed in the disinfectant for a minimum of 45 minutes. No air

bubbles should be visible on the immersed components during this soak time. The com-

ponents should then be removed from the disinfectant and immersed in purified water to

rinse off all disinfectant. Immersion should be for a minimum of one minute and the

process should be repeated two additional times using fresh purified water each time. Once

rinsing is complete, the components should be thoroughly dried using a sterile, lint-free

cloth. Note: Always allow components to completely dry before reassembling the handles,

inserting batteries, or using the devices. Lamp cartridge and bulb can be wiped with a cloth

dampened with 70% isopropyl alcohol. DO NOT ALLOW SOLUTION TO ENTER INTO THE

CARTRIDGE. The lamp cartridge and bulb CANNOT be SOAKED OR AUTOCLAVED.

Steris Sterilization Processes for Disposable and Fiberoptic Laryngoscopes: 

This product has been validated with the V-PRO 1 Standard Cycle; V-PRO 1 Plus Lumen

and Non Lumen Cycles; V-PRO maX Lumen, Non Lumen, and Flexible Cycles; V-PRO 60

Lumen and Non Lumen Cycles using the V-PRO 60 Low Temperature Sterilization System

with VAPPROX® HC Sterilant and V-PRO maX Low Temperature Sterilization System.

Steam Sterilization for Standard and Fiberoptic Blades and Handles:

Sterilizer Type:                   

Preconditioning Pulses:

Minimum Temperature:

Full Cycle Time:                

Minimum Dry Time:

Gravity

249.8°F (121°C)

30 minutes

15 minutes

Sterilizer Type:

Minimum Temperature:

Full Cycle Time:

Minimum Dry Time:

Prevacuum

3

269.6°F (132°C)

4 minutes

20 minutes

Repair/Maintenance

Battery Replacement: 

(Reuseable Models)

1)  Unscrew bottom cap of handle and remove batteries. 

2)  Replace with appropriate size battery, taking care to insert posi-

tive ends first. Screw bottom cap on and tighten. (Figure 2).

NOTE:

Alkaline batteries should be used with your handle for max-

imum performance and are recommended for replacement. Ordi-

nary carbon zinc batteries may also be used.

Main

Handle

Batteries

O  

 

 

 

 

 

Figure 2

Batteries

Main

Handle

Содержание Satin

Страница 1: ...ADC SatinTM Standard Fiber Optic and Disposable Laryngoscope Blades and Handles Directions for Use...

Страница 2: ...ications The following are only relative contraindication to tracheal intubation 1 Severe airway trauma or obstruction that does not permit safe passage of an endotra cheal tube Emergency cricothyroto...

Страница 3: ...ation of all Laryngoscope Models To Activate Attach selected blade onto handle by hooking blade onto handle rod as shown in Figure 1 Pull blade up ward to open position to automatically activate light...

Страница 4: ...aning and replacement WARNING Ultrasonic cleaning is not recommended Disinfection Sterilization Warnings for All Laryngoscope Blades CAUTION ADC laryngoscope blades do not come sterilized The Sterrad...

Страница 5: ...be wiped with a cloth dampened with 70 isopropyl alcohol DO NOT ALLOW SOLUTION TO ENTER INTO THE CARTRIDGE The lamp cartridge and bulb CANNOT be SOAKED OR AUTO CLAVED High Level Disinfection for Stan...

Страница 6: ...OLUTION TO ENTER INTO THE CARTRIDGE The lamp cartridge and bulb CANNOT be SOAKED OR AUTOCLAVED Steris Sterilization Processes for Disposable and Fiberoptic Laryngoscopes This product has been validate...

Страница 7: ...mp collar and rotate lamp counterclock wise until free 2 Replace the lamp with a new ADC lamp 3 Verify that lamp is sufficiently tightened before use CAUTION Be sure lamp s glass surface is clean and...

Страница 8: ...y reversing the above steps 4 Tighten the locking screw ensuring that the fiberoptic bundle is securely attached to the laryngoscope blade Figure 5 Test Procedures Laryngoscope blades and handles shou...

Страница 9: ...ELS DESCRIPTIONS Laryngoscope Item Mac 0 Mac 1 Mac 2 Mac 3 Mac 4 Miller 00 Miller 0 Miller 1 Miller 2 Miller 3 Miller 4 Wisconsin 0 Wisconsin 1 Wisconsin 1 5 Wisconsin 2 Wisconsin 3 Wisconsin 4 C Cell...

Страница 10: ...ll Handle Stubby Handle Fiberoptic Blade 4070F 4071F 4072F 4073F 4074F NA 4080F 4081F 4082F 4083F 4084F NA NA NA 4NA NA NA 4065F 4066F 4067F Fiberoptic Replacement Lamp NA NA NA NA NA NA NA NA NA NA N...

Страница 11: ...84D 20 NA NA NA NA NA NA NA 4065D 20 NA SPU MODELS DESCRIPTIONS REPLACEMENT PARTS CHART Laryngoscope Item Mac 0 Mac 1 Mac 2 Mac 3 Mac 4 Miller 00 Miller 0 Miller 1 Miller 2 Miller 3 Miller 4 Wisconsin...

Страница 12: ...ental special or consequential damages so this limitation may not apply to you To Obtain Warranty Service Send item s postage paid to ADC Attn Repair Dept 55 Commerce Dr Hauppauge NY 11788 Please incl...

Страница 13: ...it 6 PO14 1TH United Kingdom Inspected in the U S A Reusable models Made in Pakistan Disposable models Made in Taiwan China tel 631 273 9600 toll free 1 800 232 2670 fax 631 273 9659 www adctoday com...

Отзывы: