background image

10

EN

Never over-infl ate the cuff.

 Avoid prolonged intracuff pressures greater than 60 cm H

2

O. The initial cuff 

pressure varies according to patient, mask size, head position, and depth of anaesthesia. 
Do not hold the tube during infl ation as this prevents the mask from seating itself correctly. A small outward 
movement of the tube may be seen as the mask is infl ated.
To avoid over infl ation, it is very important to strictly adhere to the cuff-infl ation volumes stated in Table 3.
Over-infl ation can be entirely avoided by completely defl ating the cuff prior to insertion by withdrawing all of the 
air with a suitable syringe. This is the method recommended by Ambu. 
In instances where an alternative technique is adopted, for example, if the cuff is inserted in a neutral or
semi-infl ated state, there is a risk that the cuff may be over-infl ated. Once the mask is inserted extra care must 
be taken to compensate for the air already in the mask when subsequently infl ating the cuff. The maximum extra 
volume depends on mask size and initial volume of air in the mask when inserted.

WARNING

Never overinfl ate the cuff after insertion.

Look for the following signs of correct placement: The possible slight outward movement of the tube upon cuff 
infl ation, the presence of a smooth oval swelling in the neck around the thyroid and cricoid area, or no cuff visible 
in the oral cavity. 

7.6. Connecting to the Anaesthetic System

Carefully connect the Ambu AuraOnce to the anaesthetic circuit or ventilation bag and initiate gentle manual 
ventilation, looking for any signs of leakage. Auscultation over the lungs and epigastrium and capnography should 
be used to determine suffi cient respiration. Auscultate in the anterolateral neck region to check for abnormal 
sounds that might indicate mild laryngeal spasm or light anaesthesia.
The mask may leak slightly for the fi rst three or four breaths before settling into position in the pharynx. In case 
leakage persists, check that there is adequate depth of anaesthesia and that the pulmonary infl ation pressures are 
low before assuming that reinsertion of the Ambu AuraOnce is necessary. 
As with other methods of airway management, use of pulse oximetry and capnography is recommended when 
using the Ambu AuraOnce. The mask can be used for either spontaneous or controlled ventilation.

WARNING

•  Any signs of airway problems or inadequate ventilation must be monitored regularly and the Ambu AuraOnce 

must be replaced or removed as required to maintain a patent airway.

•   During anaesthesia, nitrous oxide may diffuse into the cuff causing an increase in cuff volume/pressure. 

Cuff pressure should be monitored and adjusted routinely.

•  The anaesthetic breathing system must be adequately supported when connected to the Ambu AuraOnce 

to avoid rotation of the mask.

•  The patency of the Ambu AuraOnce should be reconfi rmed after any change in the patient’s head or 

neck position

7.7. Fixation

Secure the Ambu AuraOnce to the patient’s face with adhesive tape or with a mechanical tube holder suited for 
this purpose. Do not use an oral Guedel airway as a bite block because it will prevent correct positioning of the 
mask increasing the risk of trauma and reducing seal effectiveness. It is recommended to use a gauze bite block. 

See fi gure 

i

. Fixation of Ambu AuraOnce.

In order to prevent stimulation of the patient’s airway do not reposition or move the laryngeal mask during use and avoid 
moving the patient during anaesthesia to prevent stimulation of the airway. 

Содержание AuraOnce

Страница 1: ...Product Information Ambu AuraOnce Single Use Laryngal Mask Sterile For use by trained clinicians only...

Страница 2: ...7 3 1 Placement technique 9 7 4 Insertion problems 9 7 5 Inflation 9 7 6 Connecting to the anesthetic system 10 7 7 Fixation 10 7 8 Usage with spontaneous ventilation 11 7 9 Usage with positive press...

Страница 3: ...ery equipment Use of a nasogastric tube may make regurgitation likely because the tube may interfere with the function of the lower esophageal sphincter Do not attempt to clean and reuse the Ambu Aura...

Страница 4: ...should only be used in patients who have been clinically evaluated by a clinician familiar with anesthesia as eligible for a laryngeal mask airway When the Ambu AuraOnce is used in profoundly unconsci...

Страница 5: ...1 g Internal volume of ventilatory pathway 5 5 ml 8 ml 11 ml 15 ml 16 ml 21 ml 30 ml 38 ml Pressure drop Min interdental gap 15 mm 17 mm 19 mm 21 mm 25 mm 29 mm 31 mm 32 mm i Internal pathway 10 3 cm...

Страница 6: ...shown in table 3 are maximum volumes Applying the stated maximum inflation volume may respond to a cuff pressure above the maximum of 60 cm H2 O It is recommended to continuously monitor the cuff pre...

Страница 7: ...ot be any bulge nor any sign of leakage in the cuff pilot tubing or pilot balloon WARNING Do not use the Ambu AuraOnce if there are any bulges on the cuff or if there are any signs of leakage Table 5...

Страница 8: ...ce are familiar with the warnings precautions indications and contraindications found in these Product Information The following points are extremely important Check for correct deflation and lubricat...

Страница 9: ...nal or intravenous agents and initiate manual ventilation If you cannot open the patient s mouth sufficiently to insert the mask check that the patient is adequately anesthetized Ask an assistant to p...

Страница 10: ...d to determine sufficient respiration Auscultate in the anterolateral neck region to check for abnormal sounds that might indicate mild laryngeal spasm or light anaesthesia The mask may leak slightly...

Страница 11: ...o not simply add more air This will not necessarily improve the seal pressure and may even increase the leak by adding tension to the normally soft cuff pushing it away from the larynx Instead remove...

Страница 12: ...ipment and the facility for rapid tracheal intubation are available Do not fully deflate the cuff until after its removal to avoid secretions entering the larynx and to prevent laryngospasm Alternativ...

Страница 13: ...te that with the Ambu AuraOnce as with any form of airway management and anesthesia in pediatric patients where ventilation is insufficient desaturation is likely to occur faster because of the higher...

Страница 14: ...14 EN c e 1 2 6 5 4 3 7 30 ml 60 cm H 2 O 50 7 0 kg 4 d A J I K H G D C B F E 2 6 5 4 3 1...

Страница 15: ...15 EN g i h f...

Страница 16: ...Catalog number LOT REF Copyright 2008 Ambu A S Denmark All rights reserved No part of this programme or the programme documentation may be reproduced in any form including photocopying without the pri...

Страница 17: ...ermany Ambu GmbH In der Hub 5 D 61231 Bad Nauheim Tel 49 6032 92500 Fax 49 800 ambude www ambu de UK Ambu Ltd 8 Burrel Road St Ives Cambridgeshire PE27 3LE Tel 44 0 1480 498 403 Fax 44 0 1480 498 405...

Страница 18: ...497 3600 01 11 2008 Version 1 0 1 Ambu A S Baltorpbakken 13 DK 2750 Ballerup Denmark T 45 7225 2000 F 45 7225 2050 www ambu com 0086 LATEX FREE LATEX ATE Pat Pending...

Отзывы: