background image

EN

17

12.5 Disinfection of the Ambu® aScope™ Monitor 

The Ambu® aScope™ Monitor must be cleaned and disinfected according to the instructions before first use.
Immediately after use, remove any soiling from the exterior of the Ambu® aScope™ Monitor.

1.   Wipe the surfaces of the Ambu® aScope™ Monitor during approximately 15 minutes using a piece of sterile gauze dampened with the alcohol  

mixture indicated below (approximately once every 2 minutes). Follow safety procedures for handling isopropyl. The gauze should not be dripping 
since liquid can affect the electronics inside the Ambu® aScope™ Monitor. Pay close attention to buttons, screen, external casing, slots and gaps on 

the Ambu® aScope™ Monitor. Use a sterile cotton swab to reach these areas.

Solution 

Concentration 

Preparation

Isopropyl (alcohol) 95% 

70-80% 

 80cc of 95% Isopropyl (alcohol) added to 20cc of purified water (PURW)*

 

*    Alternatively, use EPA-registered hospital desinfection wipes containing at least 70% isopropyl. Safety precautions and use directions of the  

manufacturer shall be followed.

2.   After cleaning and disinfection, the Ambu® aScope™ Monitor must be submitted to the pre-check procedure in section 11.
3.   The Ambu® aScope™ Monitor must be stored between procedures in accordance with local guidelines

12.6 Disposal

The Ambu® aScope™ is a single patient use device. The Ambu® aScope™ is considered infected after use and must be disposed of in accordance with 
local guidelines for collection of infected medical devices with electronic components. 

The Ambu® aScope™ Monitor is a reusable device. At the end of its product life, the Ambu® aScope™ Monitor and power supply must be disposed of in 
accordance with local regulations and collection schemes for disposal of electronic and electrical devices. 

The battery inside the Ambu® aScope™ Monitor is to be removed only at the end of the product lifetime. The battery must be disposed of in accordance  
with local regulations and collection schemes for disposal of batteries. In order to remove the battery, use a Torx screwdriver to remove the battery lid.

13. Trouble-shooting guide

If problems occur with the system, please use this trouble-shooting guide to identify the cause and correct the error. 

No Picture 

Cause 

Action

 Ambu® aScope™ not connected to Ambu® aScope™ Monitor 

Connect an Ambu® aScope™

 Ambu® aScope™ Monitor not turned on 

 Turn monitor on. Allow about 10 seconds start-up time.

 Brightness and contrast settings are wrong 

 Adjust contrast and brightness using the designated buttons on the monitor

Ambu® aScope™ not turned on 

Turn Ambu® aScope™ on

 No power on Ambu® aScope™ Monitor 

 Charge Ambu® aScope™ Monitor or connect the Ambu® aScope™ Monitor to the 
power supply. This can be done during the procedure

Low picture quality 

Cause 

Action

Light reflecting on the monitor screen 

 Move the monitor to a position where no direct light influences the screen

Dirty/damp screen 

 Wipe the screen with a clean cloth

 Brightness and contrast settings not optimal 

 Adjust the contrast and brightness using the designated buttons on the monitor

 Blood, saliva etc. on the lens (distal tip) 

 Wipe the lens with a clean cloth

14. Warranty and replacement program

The warranty period for the Ambu® aScope™ Monitor is one year from delivery to the customer. We agree to replace an Ambu® aScope™ Monitor free of 
charge if proof can be provided of faulty materials or faulty workmanship. In doing so we cannot accept the cost of transportation or risk of shipment. 
There is no warranty on the Ambu® aScope™.

A defective Ambu® aScope™ Monitor must be handled exclusively by persons authorised by Ambu A/S. During our inspection of the Ambu® aScope™ 
Monitor you will receive an identical replacement of the Ambu® aScope™ Monitor. 
To prevent infection, it is strictly forbidden to ship contaminated medical devices. The medical device (Ambu® aScope™ Monitor or Ambu® aScope™) 
must be decontaminated on site before shipment to Ambu. The cleaning and disinfection procedures explained in 12.4 and 12.5 shall be followed. Ambu 
reserve the right to return contaminated medical devices to the sender.

Содержание aScope

Страница 1: ...ns for use Ambu aScope For Single Patient use only Sterile For use by trained clinicians physicians only For in hospital use Ambu aScope Monitor For use by trained clinicians physicians only For in ho...

Страница 2: ...800 ambude www ambu de UK Ambu Ltd 8 Burrel Road St Ives Cambridgeshire PE27 3LE Tel 44 1480 498 403 Fax 44 1480 498 405 www ambu com Italy Ambu S R L Via Paracelso 18 Centro Direzionale Colleoni 2004...

Страница 3: ...to an external monitor video recorder 15 10 Clinical Application 15 11 Preparation for Use 15 12 Use of the Ambu aScope system 16 12 1 Injection and control of the Luer channel 17 12 2 Removal proced...

Страница 4: ...loose parts or the bending section does not work as intended Do not touch the distal tip of the insertion cord or allow it to strike other objects The lens surface of the distal tip is fragile and vis...

Страница 5: ...conventional cleaning and sterilization procedures 2 Intended Use The Ambu aScope is intended for use as an aid in the placement of an Endotracheal Tube ETT directly or through an intubating laryngea...

Страница 6: ...e Ambu aScope Working length of the Ambu aScope insertion cord Maximum width of inserted portion Min ETT ID 6 mm Minimum ET tube inner diameter 6 0mm see section 12 Field of view Operating time of the...

Страница 7: ...and electrical waste WEEE Only applicable for the Monitor Li ion Battery type Lithium ion Only applicable for the battery inside the Ambu aScope Monitor Re chargeable battery Only applicable for the...

Страница 8: ...duct specifications Ambu aScope Optical System Field of View 80 Direction of View 0 forward viewing Depth of Field 2mm 50 mm 0 1 1 968 inch Illumination method LED LTW C282DS5 SE Insertion cord Distal...

Страница 9: ...al standard NTSC Video out connection 3 5mm jack socket Video out cable 3 5mm jack plug to RCA jack plug Storage Temperature 10 40 C 50 104 F Relative humidity 10 90 Mounting interface Mounting interf...

Страница 10: ...r battery icon on the screen changes from green fully charged battery via orange half battery capacity to red low bat tery capacity within 30 minutes it is recommended that the Ambu aScope Monitor is...

Страница 11: ...t Function Material 1 Monitor frame PC Polycarbonate ABS Acrylonitrile Butadiene 2 Monitor screen Shows the picture from the camera Glass 3 Bracket Secures the monitor to e g an IV pole POM GF25 Polyo...

Страница 12: ...sportation purposes Cardboard Table 5 Functions and materials 8 2 1 Monitor placement on the bracket The Ambu aScope Monitor can be placed on a flat surface But to secure the Ambu aScope Monitor it is...

Страница 13: ...alf time left orange Time is expiring Less than 5 minutes left red Plug is connected and the monitor battery is above half charged green Plug is connected and the monitor battery is half charged orang...

Страница 14: ...device should not be used A physician who is familiar with this device can perform the preparation for use in a time range of 1 4 minutes The tests should be conducted in a manner consistent with acce...

Страница 15: ...tructions do not explain or discuss clinical intubation procedures They describe only the basic operation and precautions related to the operation of this Ambu aScope system Before initial use of the...

Страница 16: ...rocedure turn the Ambu aScope off in between sessions and place it on a sterile surface Be aware of the total operating time limit of 30 minutes during a period of 8 hours from first switching on 12 3...

Страница 17: ...d collection schemes for disposal of batteries In order to remove the battery use a Torx screwdriver to remove the battery lid 13 Trouble shooting guide If problems occur with the system please use th...

Отзывы: