background image

EN

10

Plugs

Between the power supply and 

Ø3.4mm DC jack connector 

Ambu® aScope™ Monitor 
4 interchangeable types  

 1) Class II model NEMA 1-15P AC power plug with 2 prongs 
2) Australian configuration: SAA 2 pins, class II 

3) UK configuration: UK 2 pins, class II

 

4) European configuration: Europlug 2 pins, class II

 
Table 3 - Specification for the Ambu® aScope™ power supply

8. Ambu® aScope™ functions

The Ambu® aScope™ system consists of the Ambu® aScope™ and Ambu® aScope™ Monitor. To avoid a possible risk of cross-contamination the  
Ambu® aScope™ is a sterile single patient use device. 

The Ambu® aScope™ system will prevent reuse of the Ambu® aScope™. The Ambu® aScope™ can be used/switched on for a total of 30 minutes within an 
8 hour period from first switching on. When the 30 minute or 8 hour period is exceeded the power to the Ambu® aScope™ system is switched off. When 
the Ambu® aScope™ is connected to the Ambu® aScope™ Monitor, do not disconnect the Ambu® aScope™ from the Ambu® aScope™ Monitor within the 
first two minutes.
The time remaining for the Ambu® aScope™ will be shown as symbols on the Ambu® aScope™ Monitor at all times when powered on and connected. 
See section 8.2.2 for explanation of indicators. The Ambu® aScope™ Monitor is reusable.
If the Ambu® aScope™ Monitor battery icon on the screen changes from green (fully charged battery), via orange (half battery capacity) to red (low bat-
tery capacity) within 30 minutes, it is recommended that the Ambu® aScope™ Monitor is replaced.

The Ambu® aScope™ Monitor can be used for at least 150 intubations if it is stored, used and cleaned as described in this instruction for use.

8.1 Ambu® aScope™ 

The optical section of the Ambu® aScope™ consists of a camera in a flexible and sealed distal tip. Like all optical systems these parts are very delicate. 
Therefore careful handling of the Ambu® aScope™ is recommended. 

When turning the Ambu® aScope™ off in between sessions – do not unplug. Simply press the on/off button.

1

2

3

4

5

6

7

8

9

10

Figure 1 - The Ambu® aScope™

Number 

Part 

Function 

Material

1   

Distal tip  

Camera 

Epoxy

2    

Bending section 

Manoeuvrable part 

PU (Polyurethane)

3    

Insertion cord 

Flexible airway insertion cord 

PU (Polyurethane)

4   

Connector on  

Connects to Ambu® 

PE (Poly Ethylene) +  

 

 

Ambu® aScope™ cable 

aScope™ Monitor  

PVC (Polyvinyl Chloride) 

5    

Ambu® aScope™ cable 

Transmits the picture 

PVC (Polyvinyl Chloride)

 

 

 

signal to the Ambu®aScope™ Monitor.

6   

ET-tube connection 

Fixes the ET-tube during procedures 

SEBS (Styrene Ethylene Butadiene)

7   

Handle 

Suitable for both right and 

MABS (Methyl Acrylonitrile

 

 

 

 

left handed use 

Butadiene Styrene) +  

 

 

 

 

SEBS  (Styrene Ethylene Butadiene)

8   

Luer channel 

The Luer entry fits a Luer 

SB (Styrene Butadiene)

 

 

 

entry connector and can be moved 

 

 

 

from side to side. 

Содержание aScope

Страница 1: ...ns for use Ambu aScope For Single Patient use only Sterile For use by trained clinicians physicians only For in hospital use Ambu aScope Monitor For use by trained clinicians physicians only For in ho...

Страница 2: ...800 ambude www ambu de UK Ambu Ltd 8 Burrel Road St Ives Cambridgeshire PE27 3LE Tel 44 1480 498 403 Fax 44 1480 498 405 www ambu com Italy Ambu S R L Via Paracelso 18 Centro Direzionale Colleoni 2004...

Страница 3: ...to an external monitor video recorder 15 10 Clinical Application 15 11 Preparation for Use 15 12 Use of the Ambu aScope system 16 12 1 Injection and control of the Luer channel 17 12 2 Removal proced...

Страница 4: ...loose parts or the bending section does not work as intended Do not touch the distal tip of the insertion cord or allow it to strike other objects The lens surface of the distal tip is fragile and vis...

Страница 5: ...conventional cleaning and sterilization procedures 2 Intended Use The Ambu aScope is intended for use as an aid in the placement of an Endotracheal Tube ETT directly or through an intubating laryngea...

Страница 6: ...e Ambu aScope Working length of the Ambu aScope insertion cord Maximum width of inserted portion Min ETT ID 6 mm Minimum ET tube inner diameter 6 0mm see section 12 Field of view Operating time of the...

Страница 7: ...and electrical waste WEEE Only applicable for the Monitor Li ion Battery type Lithium ion Only applicable for the battery inside the Ambu aScope Monitor Re chargeable battery Only applicable for the...

Страница 8: ...duct specifications Ambu aScope Optical System Field of View 80 Direction of View 0 forward viewing Depth of Field 2mm 50 mm 0 1 1 968 inch Illumination method LED LTW C282DS5 SE Insertion cord Distal...

Страница 9: ...al standard NTSC Video out connection 3 5mm jack socket Video out cable 3 5mm jack plug to RCA jack plug Storage Temperature 10 40 C 50 104 F Relative humidity 10 90 Mounting interface Mounting interf...

Страница 10: ...r battery icon on the screen changes from green fully charged battery via orange half battery capacity to red low bat tery capacity within 30 minutes it is recommended that the Ambu aScope Monitor is...

Страница 11: ...t Function Material 1 Monitor frame PC Polycarbonate ABS Acrylonitrile Butadiene 2 Monitor screen Shows the picture from the camera Glass 3 Bracket Secures the monitor to e g an IV pole POM GF25 Polyo...

Страница 12: ...sportation purposes Cardboard Table 5 Functions and materials 8 2 1 Monitor placement on the bracket The Ambu aScope Monitor can be placed on a flat surface But to secure the Ambu aScope Monitor it is...

Страница 13: ...alf time left orange Time is expiring Less than 5 minutes left red Plug is connected and the monitor battery is above half charged green Plug is connected and the monitor battery is half charged orang...

Страница 14: ...device should not be used A physician who is familiar with this device can perform the preparation for use in a time range of 1 4 minutes The tests should be conducted in a manner consistent with acce...

Страница 15: ...tructions do not explain or discuss clinical intubation procedures They describe only the basic operation and precautions related to the operation of this Ambu aScope system Before initial use of the...

Страница 16: ...rocedure turn the Ambu aScope off in between sessions and place it on a sterile surface Be aware of the total operating time limit of 30 minutes during a period of 8 hours from first switching on 12 3...

Страница 17: ...d collection schemes for disposal of batteries In order to remove the battery use a Torx screwdriver to remove the battery lid 13 Trouble shooting guide If problems occur with the system please use th...