background image

Spare Part

Description

Item numbers

Thickness: 4 pt

Ambu® aBox™ 2 -Visualization 
device interface kit - Grey-Empty-
Green

Contains: 
One grey and one green visualization 
device interface board (VDI), a front 
cover with a colour ring (grey and 
green), a plectrum tool, and two 
screws for the VDI.

505000530

Power cables - J (CH), K (DK), I (AUS/NZ)

Power cable - B (US, JP)

Power cables - G (UK), E/F (EU, not DK, 
CH)

505000520

505000521

505000522

1.6. Spare Part Replacement Process

This section specifies a spare part replacement process which is in compliance with IEC 62353.

Read these Instructions for Use carefully before performing any replacements on the Ambu® 
aBox™ 2.

The technician and personnel being responsible for the replacement process should fulfil the 
requirements outlined in section 1.1.

Follow the process in Figure 2 to replace spare parts in the displaying unit.

Use the Trouble 

Shooting guide in 

the IFU 

Contact Ambu, if 

you need further 

assistance

Order and receive 

relevant parts to be 

replaced.

Chapter 2:

Clean and disinfect 

the displaying unit

Chapter 4:

Replace the 

relevant part(s) of 

the displaying unit 

Chapter 5:

Inspect the 

displaying unit 

Chapter 2:

Clean and disinfect 

the displaying unit 

Follow the standard operating procedures 

required by your organization

Figure 2 Spare part replacement flow

8

Содержание aBox 2

Страница 1: ...For use by qualified biomedical technicians only Ambu aBox 2 SPARE PART REPLACEMENT MANUAL...

Страница 2: ...Ambu is a registered trademark and aScope and aBox are trademarks of Ambu A S 2797 2...

Страница 3: ...utions 4 1 3 Data Security 5 1 4 System Description 6 1 5 Spare Parts 7 1 6 Spare Part Replacement Process 8 2 How To Clean and Disinfect the Displaying Unit 9 3 Preparing for Spare Part Replacement 1...

Страница 4: ...start replacing any spare part of the displaying unit unless you have read through this document and are qualified in the following areas Knowledge experience and acquaintance with electronic repair a...

Страница 5: ...parts supplied by Ambu Do not modify the spare parts Cleaning and disinfection wipes shall be moist but not dripping to ensure no damage to internal electronics of the displaying unit If using wipes c...

Страница 6: ...System Description The displaying unit is intended to display live imaging data from Ambu visualization devices 7 8 19 18 17 16 14 15 13 12 10 11 1 2 3 4 5 6 9 1 3 2 4 Wi Fi Figure 1 System illustrati...

Страница 7: ...otated 10 Video output ports 2 x DVI D Enable connection to external monitor or external medical imaging recorder 11 Video output ports 2 x 3G SDI 12 Wi Fi antenna connector Enables connection of Wi F...

Страница 8: ...these Instructions for Use carefully before performing any replacements on the Ambu aBox 2 The technician and personnel being responsible for the replacement process should fulfil the requirements out...

Страница 9: ...following use Excessive soil should be removed on accessible areas of the device excluding electrical ports Limitations The displaying unit is not compatible with ultrasonic or automatic cleaners and...

Страница 10: ...ide of rubber covers screen external casing of the monitor with the moist gauze Avoid getting the displaying unit wet to prevent damaging internal electronic components 4 Wait for 10 minutes or the ti...

Страница 11: ...ys use ESD protection when disassembling replacing a spare part and reassembling the displaying unit To avoid malfunction of the equipment only use spare parts supplied by Ambu Do not modify the spare...

Страница 12: ...at the original positions a Detach the front cover from the displaying unit Locate the groove on one of the sides of the front cover where a snap feature is located on the inside Press inwards with t...

Страница 13: ...Remove the 2 screws holding each VDI board using a Phillips screwdriver d Pull out the VDI boards from the displaying unit and dispose them according to local guidelines for disposal of electronic was...

Страница 14: ...h the supplied screws using the Phillips screwdriver into their original positions g Attach the new front cover supplied with the replacement kit Start by attaching from one end Ensure that the snap f...

Страница 15: ...inspection with a functioning scope Connect the displaying unit to a suitable power outlet via the power cable II Connect a functioning visualization device to the displaying unit by plugging it into...

Страница 16: ...495407500 V02 2022 04 Ambu A S Baltorpbakken 13 2750 Ballerup Denmark T 45 72 25 20 00 ambu com...