e)
Carefully insert the new VDI boards into their original positions
Ensure that the VDI boards are oriented correctly as shown on the picture below.
NOTE:
Insert the VDI boards straight and horizontally into the original main board
connector positions to ensure correct connection of the VDI board to the displaying unit.
THIS SIDE UP
THIS SIDE UP
f)
Carefully fasten the VDI boards with the supplied screws using the Phillips
screwdriver into their original positions
g)
Attach the new front cover supplied with the replacement kit
Start by attaching from one end. Ensure that the snap features on each end is placed
inside the housing of the displaying unit. The front cover is pressed in place until all top
and bottom snap features click in place.
NOTE:
Ensure that the colour rings are fitted correctly in the front cover and that colours
and arrows on the front cover and the displaying unit match.
NOTE:
Ensure the USB cover flap is inserted underneath the USB port.