2. How To Clean and Disinfect the Displaying Unit

CAUTIONS

 

• 

Cleaning and disinfection wipes shall be moist, but not dripping to ensure no 
damage to internal electronics of the displaying unit. 

• 

If using wipes containing hypochlorite or citric acid, ensure that all residue is 
completely removed. Wipes containing hypochlorite or citric acid may affect the 
screen's antireflective coatings over time. You should limit the use of wipes 
containing hypochlorite or citric acid to required cases only. 

The displaying unit is a reusable medical device. According to the Spaulding classification, the 
displaying unit is a non-critical device. 
The displaying unit should be cleaned and disinfected before and after repair and replacement 
procedures following one of the procedures below.

Any deviation from the instructions should be properly evaluated for effectiveness and 
potential adverse consequences by the person responsible for cleaning and disinfection, to 
ensure that the device continues to fulfil its intended purpose. Cleaning procedures should 
begin as soon as possible following use. Excessive soil should be removed on accessible areas 
of the device, excluding electrical ports.

Limitations:

 The displaying unit is not compatible with ultrasonic or automatic cleaners and 

should not be immersed.

Procedure 1 - Cleaning and disinfection with hypochlorite

Hypochlorite-based wipes approved for disinfection of medical devices, e.g., Sani-Cloth® 
Bleach from PDI, should be used in accordance with the wipe manufacturer’s instructions.

Cleaning: 

Use a wipe to remove heavy soil. All blood and other body fluids must be 

thoroughly cleaned from surfaces and objects. Inspect the displaying unit for cleanliness, 
function and integrity before disinfection by germicidal wipe. If visible soil remains, 
reclean the displaying unit.

Disinfecting:

1. 

For heavily soiled surfaces, use a wipe to pre-clean the displaying unit prior to 
disinfecting.

2. 

Unfold a clean wipe and thoroughly wet the surface of the displaying unit.

3. 

The treated surfaces must remain visibly wet for a full four (4) minutes (or the time 
recommended by the manufacturer of the disinfectant, at least 4 minutes). Use 
additional wipes if needed to assure continuous 4 minutes wet contact time.

4. 

Let the displaying unit air dry.

Procedure 2 - Cleaning and disinfection with quaternary ammonium compounds

9

Содержание aBox 2

Страница 1: ...For use by qualified biomedical technicians only Ambu aBox 2 SPARE PART REPLACEMENT MANUAL...

Страница 2: ...Ambu is a registered trademark and aScope and aBox are trademarks of Ambu A S 2797 2...

Страница 3: ...utions 4 1 3 Data Security 5 1 4 System Description 6 1 5 Spare Parts 7 1 6 Spare Part Replacement Process 8 2 How To Clean and Disinfect the Displaying Unit 9 3 Preparing for Spare Part Replacement 1...

Страница 4: ...start replacing any spare part of the displaying unit unless you have read through this document and are qualified in the following areas Knowledge experience and acquaintance with electronic repair a...

Страница 5: ...parts supplied by Ambu Do not modify the spare parts Cleaning and disinfection wipes shall be moist but not dripping to ensure no damage to internal electronics of the displaying unit If using wipes c...

Страница 6: ...System Description The displaying unit is intended to display live imaging data from Ambu visualization devices 7 8 19 18 17 16 14 15 13 12 10 11 1 2 3 4 5 6 9 1 3 2 4 Wi Fi Figure 1 System illustrati...

Страница 7: ...otated 10 Video output ports 2 x DVI D Enable connection to external monitor or external medical imaging recorder 11 Video output ports 2 x 3G SDI 12 Wi Fi antenna connector Enables connection of Wi F...

Страница 8: ...these Instructions for Use carefully before performing any replacements on the Ambu aBox 2 The technician and personnel being responsible for the replacement process should fulfil the requirements out...

Страница 9: ...following use Excessive soil should be removed on accessible areas of the device excluding electrical ports Limitations The displaying unit is not compatible with ultrasonic or automatic cleaners and...

Страница 10: ...ide of rubber covers screen external casing of the monitor with the moist gauze Avoid getting the displaying unit wet to prevent damaging internal electronic components 4 Wait for 10 minutes or the ti...

Страница 11: ...ys use ESD protection when disassembling replacing a spare part and reassembling the displaying unit To avoid malfunction of the equipment only use spare parts supplied by Ambu Do not modify the spare...

Страница 12: ...at the original positions a Detach the front cover from the displaying unit Locate the groove on one of the sides of the front cover where a snap feature is located on the inside Press inwards with t...

Страница 13: ...Remove the 2 screws holding each VDI board using a Phillips screwdriver d Pull out the VDI boards from the displaying unit and dispose them according to local guidelines for disposal of electronic was...

Страница 14: ...h the supplied screws using the Phillips screwdriver into their original positions g Attach the new front cover supplied with the replacement kit Start by attaching from one end Ensure that the snap f...

Страница 15: ...inspection with a functioning scope Connect the displaying unit to a suitable power outlet via the power cable II Connect a functioning visualization device to the displaying unit by plugging it into...

Страница 16: ...495407500 V02 2022 04 Ambu A S Baltorpbakken 13 2750 Ballerup Denmark T 45 72 25 20 00 ambu com...

Отзывы: