2. How To Clean and Disinfect the Displaying Unit
CAUTIONS
•
Cleaning and disinfection wipes shall be moist, but not dripping to ensure no
damage to internal electronics of the displaying unit.
•
If using wipes containing hypochlorite or citric acid, ensure that all residue is
completely removed. Wipes containing hypochlorite or citric acid may affect the
screen's antireflective coatings over time. You should limit the use of wipes
containing hypochlorite or citric acid to required cases only.
The displaying unit is a reusable medical device. According to the Spaulding classification, the
displaying unit is a non-critical device.
The displaying unit should be cleaned and disinfected before and after repair and replacement
procedures following one of the procedures below.
Any deviation from the instructions should be properly evaluated for effectiveness and
potential adverse consequences by the person responsible for cleaning and disinfection, to
ensure that the device continues to fulfil its intended purpose. Cleaning procedures should
begin as soon as possible following use. Excessive soil should be removed on accessible areas
of the device, excluding electrical ports.
Limitations:
The displaying unit is not compatible with ultrasonic or automatic cleaners and
should not be immersed.
Procedure 1 - Cleaning and disinfection with hypochlorite
Hypochlorite-based wipes approved for disinfection of medical devices, e.g., Sani-Cloth®
Bleach from PDI, should be used in accordance with the wipe manufacturer’s instructions.
Cleaning:
Use a wipe to remove heavy soil. All blood and other body fluids must be
thoroughly cleaned from surfaces and objects. Inspect the displaying unit for cleanliness,
function and integrity before disinfection by germicidal wipe. If visible soil remains,
reclean the displaying unit.
Disinfecting:
1.
For heavily soiled surfaces, use a wipe to pre-clean the displaying unit prior to
disinfecting.
2.
Unfold a clean wipe and thoroughly wet the surface of the displaying unit.
3.
The treated surfaces must remain visibly wet for a full four (4) minutes (or the time
recommended by the manufacturer of the disinfectant, at least 4 minutes). Use
additional wipes if needed to assure continuous 4 minutes wet contact time.
4.
Let the displaying unit air dry.
Procedure 2 - Cleaning and disinfection with quaternary ammonium compounds
9