Procedure
d) Each single procedural water line has to be
operated as long until the decontamination
agent “BILPRON” visibly comes out as blue
solution:
e) Switch off the treatment unit (if necessary at
the main switch!)
5x
Remarks
1. You should start with the
tumbler filler in order to flood
the main lines as soon as pos-
sible. Operate the tumbler filler
several times.
2. Then continue with the
multifunctional syringes and
finally rinse the lines with the
smallest flow.
3. All instruments which are
additionally connected to the
dental unit (dental scaler,
powder jet devices etc.) have
also to be rinsed.
4. If the tank is empty, BILPRON
has to be refilled as described
in chapter 7.2
The operating agents are on the
side cover.
l
l
l
l
l
l
l
l
l
l
l
l
min. 24 h
off