Appendix
Goldeye Technical Manual
V2.2.0
86
correction. Changing this parameter or the contents of the
specified files may void ex factory calibration.
0 : No loading of any correction image. The
memory content remains uninitialized at
startup.
01..EF : The correction data is loaded from the
indicated file. In case that the following
number also exists within the flash, it is
loaded into a second page of the correction
memory. The actually data page for correction
is selected depending on the grabber's CC2
signal state.
D=<p8> : File number of the data for the 12-bits-LUT. Changing this
parameter or the contents of the specified files may void ex
factory calibration.
0 : No LUT data is loaded. The memory content
remains uninitialized at startup.
01..EF : The correction data is loaded from the
indicated file.
E=<p8> : Operation mode of the two point correction (Gain Offset
Correction). Changing this parameter may void ex factory
calibration.
0 : Deactivated. Image data is passed through
transparently.
1 : Two point correction activated. Correction
data A and B with set values J and K of the
actual correction memory page S are employed.
2 : Two point correction deactivated. Test mode:
Correction data A is output as image data.
3 : Two point correction deactivated. Test mode:
Correction data B is output as image data.
4 : One point correction: Simple subtraction of
the correction data A from the input data.
In addition the set value J is added as
offset to each pixel value.
5 : One point correction: Simple subtraction of
the correction data B from the input data.
In addition the set value K is added as
offset to each pixel value.
F=<p8> : Operation mode of the bad pixel correction.
0 : Deactivated. Image data is passed through
transparently.
1 : Activated. Correction data C is employed.
2 : Deactivated. Test mode: Data output
controlled by L=<p8>(*).
G=<p8> : Operation mode of the LUT. Changing this parameter may void
ex factory calibration.
0 : Deactivated. Image data is passed through
transparently.
1 : Activated.
2 : Deactivated. Test mode: LUT data is output
sequentially.
H=<p8> : Operation mode of the integrator / image store. This module