Advanced parameters and commands
Goldeye Technical Manual
V2.2.0
78
Example
Set the factory default (115200 Baud and activate serial interface via Camera
Link / Gigabit Ethernet).
>
s=2A[CR]
Save parameters into flash (X)
Command
X=1
stores all parameter values into a system file within the flash
memory. After a re-start of the camera they are automatically restored.
However, neither new recorded correction images, nor modified set values (
J
and
K
for all valid
S
) are stored and will therefore be lost after the camera is
switched off.
Note
This function is not available as GigE Vision feature.
GigE Feature Name
Feature Visibility
Description
SaveParameterToCameraFlash
Expert
Stores all parameter values into a
system file within the camera‘s
flash memory.
(X=<value> command)
Table 43: GigE feature: Save parameters into flash