Advanced parameters and commands
Goldeye Technical Manual
V2.2.0
75
The table below illustrates the possible values for the individual bit groups:
Example
Integration of 32 images simultaneously to the output of the actual integration
image and copy as a reference image
A
into the correction data memory
(c=1, b=6, a=1) => ccbbba =
011101
[binary] = 0x1D. As soon as the integration
is terminated, you have to wait for at least two image cycles, before the copy
procedure is finished.
>
H=1D[CR]
H 0
Operation Mode
Integrator/Image Store
F
E
D
C
B
A
9
8
7
6
5
4
3
2
1
0
-
-
-
-
-
-
-
-
d
-
c
c
b
b
b
a
ro
rw rw rw rw rw rw
Bit group
Value (hex)
Description
a
0
Output: deactivated. Data are passed through transparently.
1
Output: activated. Output of actual image store content.
b
0
Integration: deactivated. The image store content is not changed.
1
Integration: 1 image. The following image is copied into the image
store.
4
Integration: 8 images. The following 8 images are integrated and the
result is copied into the image store.
5
Integration: 16 images. The following 16 images are integrated and the
result is copied into the image store.
6
Integration: 32 images. The following 32 images are integrated and the
result is copied into the image store.
7
Integration: 64 images. The following 64 images are integrated and the
result is copied into the image store.
c
0
Copy: deactivated.
1
Copy: Target A. The image store content of the integrator is copied into
the correction data store as low reference image of the two point cor-
rection
2
Copy: Target B. The image store content of the integrator is copied into
the correction data store as high reference image of the two point cor-
rection
d
0
Integration completed. (This value can only be read out.)
1
Integration still running. (This value can only be read out.)
Note
: The Integration is activated by
change
of the register content of bit group x of
H
. In case
that e.g. directly following the integration of 64 images a further integration of 64 images is
started, another value has to be written into the register first.
Example
:
H=E
,
H=0
,
H=E
.
Table 40: Integrator and image store: values