Remote control commands
R&S
®
NGU
145
User Manual 1179.2531.02 ─ 03
CURR:PROT:TRIP?
//response: 1 OCP is tripped
//response: 0 OCP is not tripped
//resets a tripped OCP
CURR:PROT:CLEar
[SOURce:]CURRent:PROTection:CLEar
...........................................................................145
[SOURce:]CURRent:PROTection:DELay:INITial
................................................................ 145
[SOURce:]CURRent:PROTection:DELay
.......................................................................... 146
[SOURce:]CURRent:PROTection:TRIPped?
..................................................................... 146
[SOURce:]CURRent:PROTection[:STATe]
.........................................................................147
................................................................................................ 148
[SOURce:]CURRent:PROTection:CLEar
[<Channel list>]
Resets the OCP state of the selected channel. If an OCP event has occurred before,
the reset also erases the message on the display.
Setting parameters:
<Channel list>
<list>
Example:
CURR:PROT:CLE (@1)
Resets OCP state at channel 1.
Example:
See
Example "Configuring the overcurrent protection"
Usage:
Setting only
[SOURce:]CURRent:PROTection:DELay:INITial
<duration>
[SOURce:]CURRent:PROTection:DELay:INITial?
[<Channel list>]
Sets or queries the initial fuse delay time once output turns on.
Parameters:
<duration>
<numeric value> | MIN | MINimum | MAX | MAXimum | <list>
<numeric value>
Numeric value for initial fuse delay.
MIN | MINimum
Min value for initial fuse delay.
MAX | MAXimum
Max value for initial fuse delay.
Range:
0.00 to 60.00
*RST:
0
Default unit: s
Parameters for setting and query:
<Channel list>
<list>