Remote control commands
R&S
®
NGU
147
User Manual 1179.2531.02 ─ 03
[SOURce:]CURRent:PROTection[:STATe]
<arg0>[, <Channel list>]
[SOURce:]CURRent:PROTection[:STATe]?
[<Channel list>]
Sets or queries the OCP state.
Parameters:
<arg0>
1
Activates the OCP state.
0
deactivates the OCP state.
Parameters for setting and query:
<Channel list>
<list>
Example:
CURR:PROT 1
Activates the OCP.
Example:
CURR:PROT? (@1)
Queries OCP state at channel 1.
Example:
See
Example "Configuring the overcurrent protection"
FUSE:DELay:INITial
<delay>
FUSE:DELay:INITial?
[<Channel list>]
Sets or queries initial delay time for the fuse to take effect.
Parameters:
<delay>
<numeric value> | MIN | MINimum | MAX | MAXimum | <list>
<numeric value>
Numeric value for initial fuse delay.
MIN | MINimum
Min value for lowe initial fuse delay.
MAX | MAXimum
Max value for initial fuse delay.
Range:
0.00 to 60.00
*RST:
0
Default unit: s
Parameters for setting and query:
<Channel list>
<list>
Example:
FUSE:DEL:INIT? (@1)
Queries initial fuse delay time at channel 1.
Example:
Example "Configuring the over-
Configuration commands
www.allice.de
Allice Messtechnik GmbH