Remote control commands
R&S
®
NGU
143
User Manual 1179.2531.02 ─ 03
MIN | MINimum
Minimum value of the modulation gain.
MAX | MAXimum
Maximum value of the modulation gain
UP
Increases gain by a defined step size.
DOWN
Decreases gain by a defined step size.
<arg1>
<list>
Example:
MOD:GAIN?
Return modulation gain.
7.5.10
Power line cycle setting
[SENSe:]NPLCycles
<arg0>[, <arg1>]
[SENSe:]NPLCycles?
[<arg1>]
Sets or queries the number of power line cycles for measurements.
Parameters:
<arg0>
Number of power line cycles for measurements.
Parameters for setting and query:
<arg1>
<list>
Example:
NPLC 1
Sets number of power line cycles to 1.
7.5.11
OCP setting
The
CURRent:PROTection
subsystem contains the commands for overcurrent pro-
tection parameters such as activating fuses and setting fuse parameters of the output
channels. The default unit is A.
The delay function of the fuses takes effect when the corresponding channel is activa-
ted (Output On).
Example: Configuring the overcurrent protection
This example contains all commands to configure and query the fuse states and set-
tings.
// ************************************************
// Configuring the overcurrent protection
// ************************************************
// activates the overcurrent protection
CURR:PROT 1
// queries the state of the overcurrent protection in the selected channel
Configuration commands