13
ejected from the Ultravision generator by pulling the ejection trigger toward the front of the generator.
Caution
The battery or Mains Converter must be inserted in the correct orientation. If not, damage to the battery, Mains Converter
and/or generator may result and create a fire hazard.
Risk of fire, explosion or burns. Do not heat batteries above 60°C.
Ultravision™ generator – Rear view
1 – Volume control
1 – Volume control for audible indicators
The Ultravision generator has four audible indicators that accompany the visual indicators:
Indicator
Tone
Description
Frequency
Visual indicator
Low battery
High (1047 Hz)
A single 1 second tone
2 seconds
Low battery LED
Proximity
Medium (523.3 Hz)
A single 0.3 second tone
6 seconds
Proximity LEDs
Generator fault
Low (261.6 Hz)
A single 1 second tone
2 seconds
Fault (!) LED
Power ON reminder
Low (261.6 Hz)
A single 1 second tone
15 minutes
None
The generator sounds the three different tones in sequence when switched on. The volume of these indicators may be
adjusted by twisting the volume control on the rear of the generator. The volume is increased by turning the volume control
clockwise, and reduced by turning anti-clockwise.
Warning
Ensure that the audible indicator can be heard within the operating theatre environment, and that the output LED indicators
are visible to the operating personnel. Failure to observe indications so may result in:
• Capacitive charging of electrosurgical instruments and the risk of atrial fibrillation.
• Loss of function caused by power failure or a fault within the generator.
Ultravision™ Battery Recharging Station Option – Top view
1 – Power and fault indicator
2 – Battery docking bay 1 and bay 2