IV. 10. Troubleshooting
9-GB
Observations
Probable causes
Solutions
GB
European Representative:
Gavin Ayling
9 Bungham Lane
Penkridge Stafford
Staffordshire ST19 5NH England
E-mail: [email protected]
The 0-I button is in the ON position.
The device does not operate. The
intermittent single alarm sounds.
Oxygen concentration indicator remains
lit yellow.
The alarm test does not work.
The 0-I button is ON and the
compressor is operating but there is
no flow. The audible alarm sounds.
The 0-I button is ON, the compressor
is operating, there is a flow but the
audible alarm sounds.
The compressor stops in mid-cycle,
then starts again after a few minutes.
The oxygen air flow is interrupted at
the nasal cannula outlet.
The flow at the nasal cannula outlet
is irregular.
Power cable not plugged in correctly.
Mains power failure.
Oxygen concentration is too low.
Internal electrical fault.
Pneumatic connection broken or
compressor problem.
Internal electrical fault.
Pneumatic circuit fault.
Compressor thermal safety device has
been tripped. Fan not working.
Tube disconnected or humidifier not
tight.
Pneumatic circuit problem.
Check the cable connection.
Reset the circuit breaker (2) if
necessary by pressing. Check the fuses
or circuit breaker fitted on the premises.
Contact your distributor.
Contact your distributor.
Note: if the unit has not been used for
an extended time period, it needs to
operate for several minutes before the
mains power cut alarm can become
activated.
Stop the device by pressing the 0-I
button and contact your distributor.
Stop the device and contact your
distributor.
Stop the device and wait for it to cool
down. Check that the patient circuit is
not obstructed. Start up again. If the
device does not start, contact your
distributor.
Check the gas administration circuit.
Contact your distributor.
Содержание visionaire
Страница 2: ......
Страница 4: ...AirSep Corporation VisionAire MN137 1 Rev B 11 09 V5001001 Fig I 1 Fig I 2 FR Fig I 3...
Страница 14: ......
Страница 16: ...VisionAire AirSep Corporation MN137 1 Rev B 11 09 V5001001 Abb I 1 Abb I 2 DE Abb I 3...
Страница 26: ......
Страница 28: ...VisionAire AirSep Corporation MN137 1 Rev B 11 09 V5001001 Figura I 1 Figura I 2 ES Figura I 3...
Страница 38: ......
Страница 40: ...VisionAire AirSep Corporation MN137 1 Rev B 11 09 V5001001 I 1 I 2 EL I 3...
Страница 43: ...VisionAire F AirSep Corporation MN137 1 Rev B 11 09 II II 1 0 VisionAire 15 50 I 5 5 LPM LPM 90 5 0 1 3 EL EL...
Страница 44: ...VisionAire AirSep Corporation MN137 1 Rev B 11 09 III III 1 VisionAire III 2 4 EL EL...
Страница 50: ......
Страница 52: ...VisionAire AirSep Corporation MN137 1 Rev B 11 09 V5001001 Fig I 1 Fig I 2 GB Fig I 3...
Страница 62: ......
Страница 64: ...VisionAire AirSep Corporation MN137 1 Rev B 11 09 V5001001 Fig I 1 Fig I 2 IT Fig I 3...
Страница 74: ......
Страница 76: ...VisionAire AirSep Corporation MN137 1 Rev B 11 09 V5001001 Afb I 1 Afb I 2 NL Afb I 3...
Страница 86: ......
Страница 88: ...VisionAire AirSep Corporation MN137 1 Rev B 11 09 V5001001 Fig I 1 Fig I 2 PT Fig I 3...
Страница 98: ......
Страница 100: ...VisionAire AirSep Corporation MN137 1 Rev B 11 09 V5001001 Fig I 1 Fig I 2 SV Fig I 3...
Страница 110: ......
Страница 112: ...VisionAire AirSep Corporation MN137 1 Rev B 11 09 V5001001 Fig I 1 Fig I 2 DA Fig I 3...
Страница 122: ...MN137 1 Rev B 11 09...