Operating Instructions
|
airLPT216
Version 1.2
|
Dated: 07.03.2017
LEAk DETEcTION MODE
Leak detection mode
This mode is used to apply a constant positive pres-
sure to the duct system for a specific period of time,
to search for possible leaks during this time.
The air hose must be connected to the positive pres-
sure connection.
No volumetric flow measurement
or display takes place.
After selecting this mode, the saved parameters from
the previous use of Leak Detection mode are shown
on the display, namely
•
in line 1
the test pressure
•
in line 2
the duration
The buttons have the following labelling and meaning:
F1 - MAIN
Back to Main menu
F2 - chANGE
Switch to the submenus to change the
test time and duration parameters. An
arrow in the respective line indicates the
menu to which the program switches.
The arrow can be moved vertically using
the arrow buttons and .
F3 - sTART
Start Leak Detection mode.
In the TEsT PREss. submenu
the arrow buttons and can be used to set the
test pressure within the limits 50 to 500 Pa in 5 Pa
increments (the default pressure the first time this
mode is called up is 250 Pa). If the arrow buttons
are pressed for longer than 1.5 s, the pressure value
changes quickly.
F2 – chANGE
Change the “Pressure” parameter
In the DURATION submenu
the arrow buttons and can be used to set a test
duration of 1 to 30 min in 10 sec. intervals (the default
time when this mode is first called up is 15 min). If the
arrow buttons are pressed for longer than 1.5 s, the
time value changes quickly.
F2 – chANGE
Change the “Time” parameter
Leak detection mode starts after the
sTART button
(F3)
is pressed. The current pressure in Pa and the
remaining time are displayed continuously. The de-
vice switches off the fan
(compressor) after the
time has expired. The
sTOP (F3) button
can
also be used to end this
mode beforehand.
AUTOMATIC TEST
MANUAL TEST
→ LEAK DETECTION MODE
[ENTER]
F1
F2
F3
→ TEST PRESS.
250 Pa
DURATION
15:00 min
[MAIN | CHANGE | START]
F1
F2
F3
TEST PRESS.
250 Pa
[ENTER]
F1
F2
F3
DURATION
15:00 min
[ENTER]
F1
F2
F3
TEST PRESS.
158 Pa
TIME LEFT
14:55 min
[STOP]
F1
F2
F3