TEchNIcAL DATA
Technical data
•
4-line illuminated display
•
Power supply: 230 V / 50 Hz
•
Power consumption: max. 1,900 W
•
Weight: approx. 9.5 kg
•
Dimensions (L/W/H): 440 mm x 370 mm x 253 mm
•
Working range: +5 to +50 °C
•
Storage temperature: -20 to +50 °C
•
Working range: 0.08 - 36 l/s; 0.29 - 130 m
3
/h
•
Air tightness classes to EN 1507: Class A / Class B /
Class C / Class D
•
Air tightness Class Z
•
max. test pressure: -3,500 Pa / + 6,500 Pa
•
Pressure measurement
•
Ceramic measuring cell for differential pressure
measurements for -3,500 to +6,500 Pa
•
Resolution 1 Pa
•
Max. total error 1.5 % of the FS
•
Volume flow measurement
•
Mass flow rate measurement according to the calo-
rimetric measuring principle
•
within the measurement range 0.08 to 36 l/s
•
Resolution 0.01 l/s
Equipment / scope of supply
•
1 airLPT216 air leak proof tester integrated in the pla-
stic case
•
1 plastic transport case for accessories
•
1 mains cable 230 V AC (3 m long)
•
1 air hose (5 m long) with couplings for connecti-
on to the tester and to the air duct system to be
tested
•
1 hose (10 m long) for pressure measurement, with
couplings for connection to tester and to the air
duct system to be tested
•
2 spare paper rolls for record printer
•
1 operating instructions/calibration record
•