background image

 

© Copyright AirBench Ltd 2016 

 

Page 16 of 20 

ACCESSORIES 

Some accessories are only supplied if specified with purchase. 

Filter Pressure Gauges 

A filter pressure gauge is fitted as standard to the front of each VertEx. This indicates performance of the unit 
as a whole, and filter condition. Certain models show a red zone on the filter gauge; when the gauge indicates 
within this zone, filters may be blocked or require cleaning. Turn the VertEx off and inspect the filters. 

Where no red zone is shown, a filter inspection pressure drop will be recorded on the Commissioning 
Certificate. When pressure loss is greater than that shown on the Commissioning Certificate, follow the above 
procedure. 

A high pressure reading may also indicate a blocked surface. Inspect the surface for blockages and clean if 
required. Do not operate VertEx continuously with more than 75% of the face blocked. 

Hours Run Meter 

The meter is not resettable and operates at all times when the speed control or switch is powered. 

Wheels 

Standard industrial rubber wheeled swivel castors are used. The two front wheels are lockable. Lock by 
pressing on the lever and trip by flicking (models and operation vary slightly). Do not attempt to move while 
brakes are on. 

Compressed Air Interlink 

This option ensures the safety of operators by preventing the use of air tools without extraction being 
operational; and reduces energy use by switching off the fan after 5 minutes of non-use. 

4 off series 19 airline connectors are provided to the front or side of the unit and an internal timer fitted, pre-set 
to an agreed value 

 usually 5 minutes. The control panel is supplied with an additional Red / Green push-

button. 

System control will be as follows: 

Switch main on/off switch to On 

 system is energized. Airline is switched to operational. 

Press green button 

 fan starts and runs for 5 minutes. 

Use airline or press green button again 

 timer is reset for a further 5 minutes. 

Press red button 

 fan stops, airline stops. 

The timer can be re-set to a different interval; contact the manufacturer for details.

 

 

Содержание VertEx VA

Страница 1: ...Copyright AirBench Ltd 2016 Page 1 of 20 VertEx VA Operation and Maintenance This Edition 19 November 2019 AirBench Ltd 6b Commerce Way Colchester CO2 8HR www airbench com 44 0 1206 791191 AIRBENCH LT...

Страница 2: ...ACCESSORIES 6 Commissioning 7 GENERAL 7 IN HOUSE COMMISSIONING 7 Operation 8 PRIOR TO USE 8 GENERAL USE 8 Maintenance 10 FILTER CARE 10 SPARES 13 OTHER ROUTINE MAINTENANCE 15 Accessories 16 FILTER PR...

Страница 3: ...mmissioning certificate The filters are not self cleaning They must be maintained cleaned replaced as described within Do not mix incompatible materials e g steel and aluminium It is the users respons...

Страница 4: ...er specific maintenance information About VertEx VertEx is a self contained cross draught extraction system which utilises a fan and filter combination selected for your specific application to extrac...

Страница 5: ...mplete VertEx unit If using a forklift to position the top section on the base section always ensure operators are fully trained and that other users are outside the working zone of the forklift Once...

Страница 6: ...here via ductwork a discharge spigot will be visible These units are designed to connect to ductwork of the same size as the discharge spigot generally 250 or 300 mm diameter Ducting should be install...

Страница 7: ...Ensure all operators are aware of the filter maintenance routine required for this machine Once complete you should note this on the commissioning certificate and retain the certificate for future re...

Страница 8: ...full power Check the filter pressure gauge on the front of the unit If the needle on the gauge is showing in the upper red zone check the filter condition they may require cleaning or replacement This...

Страница 9: ...Copyright AirBench Ltd 2016 Page 9 of 20...

Страница 10: ...n the accompanying log sheet Take precautions and wear appropriate PPE when handing filters They may contain hazardous dusts i If using a vacuum cleaner to empty filters excessive cleaning will reduce...

Страница 11: ...Copyright AirBench Ltd 2016 Page 11 of 20 Doors can be heavy Wear appropriate hand and foot protection...

Страница 12: ...side or damaged MINIPLEAT FILTERS Mini pleat filters have fine paper pleats often around 5mm between pleats in a metal case This filter is not cleanable Replace the filter when it is stained on the c...

Страница 13: ...MEDIA PADS Media pads are green and white fibreglass pads held in a metal frame These filters are not cleanable When replacing the frame and clip which hold the pad can be reused Always replace with w...

Страница 14: ...Copyright AirBench Ltd 2016 Page 14 of 20...

Страница 15: ...tasks suggests additional issues contact the manufacturer for advice FILTER SEALS These sit behind the filters and prevent dust passing into the body of the bench and the exhaust airstream It is impor...

Страница 16: ...d operates at all times when the speed control or switch is powered Wheels Standard industrial rubber wheeled swivel castors are used The two front wheels are lockable Lock by pressing on the lever an...

Страница 17: ...bed maintenance routines are being completed and recorded within the system logbook Condition check of fan filters electrics Performance testing according to recommended face velocities Visual check t...

Страница 18: ...Copyright AirBench Ltd 2016 Page 18 of 20 TECHNICAL Arrangement...

Страница 19: ...0 700 600 175 0 48 2 4 VA121218 1270 1800 700 1200 270 0 74 3 9 VA180618 1870 1800 700 600 225 0 74 3 9 VA120624 1270 2400 700 600 190 0 48 2 4 VA121224 1270 2400 700 1200 285 0 74 3 9 VA180624 1870 2...

Страница 20: ...chinery Directive 2006 42 EC Low voltage Directive 2014 35 EU EN 60204 1 2018 Safety of machinery electrical equipment of machines general requirements EN 60335 2 80 Safety requirements for electric f...

Отзывы: