If necessary, correct the X-ray positioning
beam manually.
❯
Once the patient has been correctly positioned
using the positioning beams, deactivate the
positioning beams using the
button.
Adjusting the head supports
❯
Adjust the height of the head supports.
❯
Carefully press the head supports by hand
towards the head in order to check if they are
in the right position. The device or the head
supports are not damaged in the process.
Ideally, the head supports should make con-
tact slightly above the eye brows; correct the
position as required.
❯
Close the head supports with the
button.
To do this, just press the button briefly – don't
press and hold.
The head supports automatically touch against
the head of the patient at a defined pressure.
7.5
Start a test run
The test run ensures that the unit can perform
the image acquisition without any problems. This
prevents unnecessary exposure of the patient to
radiation.
No radiation is generated during the test
run.
Requirements:
ü
The patient is correctly positioned in the unit
using the positioning aids and the positioning
beams.
ü
The required imaging program has been
selected.
❯
Touch
Test run
on the touch screen.
❯
Tap
Run
and hold.
While doing this, constantly monitor the move-
ments of the unit. If the unit is obstructed in its
movements, release the
Run
button. The unit
stops immediately. Reposition the patient.
❯
Press
Return run
to perform the return run.
Usage
36
2210200496L41 1911V006
EN-
US