Error
Possible cause
Remedy
Error messages during start of
the acquisition of an X-ray
image or during shut down of
the PC
Energy-saving options incor-
rectly configured
❯
Deactivate the energy-saving
options in Windows and the
BIOS completely.
Supply voltage for graphic card
inadequate or incorrectly con-
nected
❯
Check the plug connections.
❯
Compare requirements of the
graphic card with the power
supply of the PC, use larger
power supply if necessary.
PC and/or graphic card fail to
comply with the specified sys-
tem requirements
❯
Set up the system in accord-
ance with the system require-
ments.
User account control (UAC) has
not been correctly configured
❯
Make the user account con-
trol settings according to the
information in the installation
instructions.
USB dongle not detected
❯
Check whether the USB don-
gle (included in scope of deliv-
ery) is plugged into the recon-
struction computer, or check
that the USB dongle is cor-
rectly plugged in.
Virus scanner prevents the
acquisition of an X-ray image
❯
Add the installation paths of
the imaging software as
exceptions in the virus scan-
ner.
Firmware of the device does not
match the software version
❯
Check software versions and
update if necessary.
The device calibration has not
been imported or incompletely
imported
❯
Carry out initial commission-
ing using the service tool
again/initially.
Door contact is not closed
❯
Check door contact and plug-
in connections of the door
contact, close door properly.
Troubleshooting
2210200496L41 1911V006
49
EN-
US