X-ray emitter
Nominal voltage, high-voltage generator
kV
50 - 99 (±10 %)
Nominal current, high voltage generator
mA
4 - 16 (± 10%, max. 75 kV 16 mA,
max. 99 kV 10 mA)
Cooling, high-voltage generator
Automatic monitoring
Shut-off at ≥ 60°C
Additional filtering at 50 kV
mm Al
in Al
2.0 + 3.0 (added automatically for CBCT)
0.08 + 0.12 (added automatically for
CBCT)
Integrated filtering at 50 kV
mm Al
in Al
0.8
0.03
Total filtering at 50 kV
mm Al
in Al
2.8 + 3.0 (added automatically for CBCT)
0.11 + 0.12 (added automatically for
CBCT)
X-ray tube model
Toshiba D-052SB
Focal spot size as per IEC 60336 X-ray
tube
mm
in
0.5
0.02
Anode angle
°
5
Anode heat capacity
kJ
35
Pulse/break ratio
1:60 or more
Duration of radiation exposure
s
0.5 - 20
Maximum current-time product per hour
mAs
960 (at 75 kV/16 mA)
4.1
X-ray tube performance data
– Maximum deviation of the peak voltage from the displayed value ± 10%
– Maximum deviation of the tube current from the displayed value ± 20%
– Maximum deviation of the exposure time from the displayed value ± 10%
– The device complies with the standards IEC 60601-1, IEC 60601-1-3 and IEC 60601-2-63.
– The lowest possible load factor is obtained with a combination of the settings 60 kV and 4 mA.
Product description
2210200496L41 1911V006
11
EN-
US