Manufacturer’s Guidance and Declaration -
Electromagnetic Immunity
The ScanX
is intended to be used within the electromagnetic environment specified below. The user of
the ScanX should assure that it is used in such an environment.
Immunity Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic
Environment - Guidance
Electrostatic
Discharge
(ESD)
IEC 61000-4-2
± 6kV contact
± 8kV air
± 6kV
± 8kV
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic mate
-
rial, the relative humidity should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for power supply
lines
± 1 kV for Input/output
lines
± 2kV
± 1kV
Mains power quality should be that of typical
commercial or hospital environment.
Surge
IEC 61000-4-5
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
± 1kV
N/A
Mains power quality should be that of typical
commercial or hospital environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5% Ur (>95% dip in
Ur) for 0.5 cycle
40% Ur (60% dip in Ur)
for 5 cycles
70% Ur (30% dip in
Ur) for 25 cycles
<5% Ur (>95% dip in
Ur) for 5 s
<5% Ur - 10ms
40% Ur - 100ms
70% Ur - 500ms
<5% Ur - 5 s
Mains power quality should be that of typical
commercial or hospital environment.
If the user of the ScanX requires continuous
operation during power mains interruptions in
the main power supply, it is recommended that
the ScanX be powered from an uninterruptible
power supply providing emergency power.
Power frequency
(50/60Hz)
magnetic field
IEC 61000-4-8
3 A/m
3A/m
Power frequency magnetic fields should be at
levels characteristic of a typical location within
a typical commercial or hospital environment.
Note:
Ur is the a. c. mains voltage prior to application of the test level.