Troubleshooting
15.4 Error messages on the touch screen.
Problem
Probable cause
Solution
Error code 1008
Connection interrupted
•
Update the firmware.
Error code 1010
Temperature of unit too high
• Allow the unit to cool down.
• Contact technician.
Error code 1022
Subassembly not initialized
• Fault in software, update the software if required.
• Contact technician.
Error code 1024
Internal data communication
fault
•
Switch the unit off and back on again.
•
Update the firmware.
• Darken the room.
• Turn the unit so that no light can fall directly into
the insertion slot.
Error code 1026
Incorrect acquisition mode
•
Select a different acquisition mode
• Inform a service technician.
•
Update the firmware.
• Reset the scanning modes to the factory settings
via the unit interface or the Imaging Software.
Error code 1100
Permitted time for scan process
exceeded
• Contact technician.
• Check the belt drive.
• Check for blockage, remove PSP from unit.
Error code 1104
Erasure unit fault
• Contact technician.
• Replace the erasure unit.
Error code 1153
Unit fault
•
Switch the unit off and on again.
•
Update firmware.
Error code 1154
Internal data communication
fault
•
Switch the unit off and on again.
•
Update firmware.
Error code 1160
Final pentaprism assembly rota-
tion
speed not reached
• Contact technician.
•
Update firmware.
• Replace the pentaprism assembly if the problem
occurs regularly.
Error code 1170
SOL sensor timeout
Fault on the laser, SOL sensor or
pentaprism assembly
• Contact technician.
•
Update firmware.
Error code 1172
SOL sensor timeout
Fault on the laser, SOL sensor or
pentaprism assembly
• Inform a service technician.
•
Update the firmware.