Air Techniques, Inc.
Page
12
Mounting
6 General Notes
6�1 Compressed air system requirements
The compressed air network to which the compressed
air system is connected must be set up so that there is a
maximum pressure of 145 psi (10 bar)�
6�2 Set up equipment
• Before installation ensure that the compressed air
available is suitable to satisfy all requirements for its
particular usage�
• Unit classification and evaluation of conformity must
be carried out on installation by the manufacturer of
the end product�
7 Transport
WARNING
During transport of the compressed air system
parts can come loose�
• Only transport the compressed air system with
no pressure in tank�
• Before transport bleed off all compressed air
from the pressure tank and the pressure hoses�
8 Set-up
8�1 Set-up location
Set-up location requirements:
- dry and well ventilated room
(for required ventilation see 4 Technical data)
- ambient temperature range of 50 and 104° F�
(+10 and +40 °C)
- max� 70 % room air relative humidity
• Installation in purpose built rooms, e�g� boiler room, must
be checked against building regulations beforehand�
• Do not operate the units in damp or wet environments�
• Clearly indicate that the room may only be entered by
staff wearing ear protectors�
Some 70% of the electrical energy taken up by the com-
pressed air station is converted into heat and given off
to the surroundings�
• If it is possible that the required room temperatures will
be exceeded then ventilation or some form of cooling
will be required and must be installed�
• The compressor module can be set up either left or
right of the control module�
• Maintain a distance between the modules of approxi-
mately 20 inches (50 cm) to allow sufficient space for
maintenance�
8�2 Module layout
Fig� 6: Distance between control and compressor mod-
ule
8�3 Set up and secure modules
The mounting materials are supplied in the scope
of delivery�
• Loosen the modules from the pallets (undo transporta-
tion safety devices)�
• Use a forklift truck or lifting equipment to move both
modules to the location chosen for set up�
• Drill required holes into the floors for the fixtures�
• Insert plugs�
• Secure the modules to the floor with bolts�
50 cm
(20 in)
50 cm
(20 in)